Background: Detection of IgG subclasses (IgGSc) is vital for the diagnosis and management of disease, especially IgG4-related diseases (IgG4-RD). This study aimed to evaluate the performances of the chemiluminescent immunoassay (CLIA) for detecting IgGSc and diagnosing IgG4-RD by IgGSc.

Methods: A total of 40 individuals with IgG4-RD, 40 with primary Sjogren's syndrome (pSS), and 40 healthy controls (HCs) were enrolled. Serum samples were collected for the simultaneous detection of IgG1, IgG2, IgG3, and IgG4 by the Siemens immunonephelometric assay and the CLIA. The correlation analysis was performed, and diagnostic value was analyzed by the receiver operating characteristic (ROC) curve.

Results: Patients with IgG4-RD had higher IgG4 (p < 0.001) and lower IgG1 (p < 0.001) than those with pSS, and HC. The results by the Siemens immunonephelometric assay and the CLIA showed a strong correlation in detecting IgG1, IgG2, IgG3, and IgG4 (r = 0.937, r = 0.847, r = 0.871, r = 0.990, all p < 0.001, respectively). The sum of IgG1, IgG2, IgG3, and IgG4 using two assays strongly correlated with total IgG by the IMMAGE 800 (r = 0.866, r = 0.811, both p < 0.001, respectively). For discriminating IgG4-RD from pSS and HC, no significant differences were observed in CLIA IgG4 and Siemens immunonephelometric assay IgG4 (z = 0.138, p = 0.891), which provided the area under the curves (AUCs) of 0.951 (p < 0.001) and 0.950 (p < 0.001), respectively. The AUCs of CLIA IgG1 and Siemens immunonephelometric assay IgG1 in distinguishing pSS from IgG4-RD and HC were 0.761 (p < 0.001) and 0.765 (p < 0.001), respectively, with no significant differences (z = 0.228, p = 0.820).

Conclusions: The CLIA and the Siemens immunonephelometric assay appeared to have good consistency with comparable diagnostic value in detecting IgGSc, especially IgG4, and IgG1 that can accurately identify IgG4-RD or pSS in clinical practice.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11073813PMC
http://dx.doi.org/10.1002/jcla.25033DOI Listing

Publication Analysis

Top Keywords

immunonephelometric assay
8
chemiluminescent immunoassay
8
detection igg
8
igg subclasses
8
clinical performance
4
performance immunonephelometric
4
assay chemiluminescent
4
immunoassay detection
4
subclasses chinese
4
chinese background
4

Similar Publications

Biochemical blood parameters may serve as biomarkers of environmental pollution, with somatoform disorder (SD) being of interest. This study examined serum glycosylated hemoglobin (HbA1C), insulin, total protein, and SD prevalence in populations from polluted areas in Kazakhstan. It assessed the predictive value of these parameters for SD, along with other risk factors.

View Article and Find Full Text PDF

Objective: Acute retinal necrosis (ARN) caused by varicella-zoster virus (VZV) is associated with changes in specific proteins in the eye's fluid, particularly matrix metalloproteinase-3 (MMP-3), an enzyme that breaks down tissue structures, and tissue inhibitor of metalloproteinase-1 (TIMP-1), which regulates MMP activity. This study aims to investigate how these proteins correlate with the progression of ARN.

Methods: We analyzed aqueous humor samples from 33 patients with ARN and 23 control patients with virus-negative uveitis.

View Article and Find Full Text PDF

Preparation and Characterization of C-Reactive Protein Dual-Particle Latex-Enhanced Immunoturbidimetric Reagents.

BME Front

December 2024

Department of Laboratory Medicine and Center of Infectious Diseases and Pathogen Biology, The First Hospital of Jilin University, Changchun 130021, China.

This study aims to couple C-reactive protein (CRP) antibodies onto latex spheres of 2 different sizes to enhance the accuracy and sensitivity of CRP detection. Furthermore, it seeks to establish a robust methodological framework crucial for advancing the development of latex-enhanced immunoturbidimetric detection reagents. CRP, an acute-phase protein, rapidly elevates in response to infections or tissue damage.

View Article and Find Full Text PDF

[Analysis of genetic association between exon polymorphisms of and and risky drinking behavior of alcoholic liver disease].

Zhonghua Yu Fang Yi Xue Za Zhi

December 2024

Shenzhen Key Laboratory of Medical Laboratory and Molecular Diagnosis,Shenzhen518035,China Department of Laboratory Medicine, The Second People's Hospital of Shenzhen,Shenzhen518035,China.

To analyze the correlation of exon rs1126671 and exon rs971074 polymorphisms with risky drinking behaviors and alcoholic liver disease. The patients with alcoholic liver disease diagnosed in the Gastroenterology Department of the People's Hospital of Hechi from November 2021 to June 2022, including 52 cases of alcoholic liver disease with positive risky drinking behaviors, 103 cases of non-alcoholic liver disease with positive risky drinking behaviors of the same gender and age, and 105 healthy subjects with no risky drinking behaviors as control groups were retrospectively analyzed. The serum total protein and albumin are detected by immunoturbidimetry and globulin is calculated by the difference method; the serum total bilirubin and direct bilirubin are detected by the nitrite oxidation method and indirect bilirubin is calculated by the difference method; alkaline phosphatase, aspartate aminotransferase, alanine aminotransferase and γ-glutamyl transferase are detected by the substrate method.

View Article and Find Full Text PDF

Evaluation of a new soluble transferrin receptor assay and comparison to three measurement procedures.

Clin Biochem

January 2025

Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, Canada; Department of Clinical Biochemistry, Laboratory Medicine Program, University Health Network (UHN), Toronto Ontario, Canada. Electronic address:

Background: Soluble transferrin receptor (sTfR) is a useful marker in the differentiation of anemia. Clinical utility is limited by lack of standardization between measurement procedures and interpretative recommendations. Our objective was to evaluate the analytical performance of a research sTfR immunoturbidimetric assay (Alinity c, Abbott Diagnostics) and compare it to three established measurement procedures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!