The safety and effectiveness of the GORE VIABAHN Endoprosthesis for treatment of symptomatic patients with peripheral artery disease (PAD) and complex femoropopliteal (FP) lesions was assessed in a real-world Japanese practice setting. : A prospective, multicenter, postmarket surveillance study was conducted from 2016 to 2017 at 64 sites in Japan. Symptomatic patients with PAD and FP lesions ⩾ 10 cm and reference vessel diameters ranging from 4.0 to 7.5 mm were eligible for enrollment. Outcome measures evaluated at 5 years were primary patency (PP), primary-assisted patency (PAP), secondary patency (SP), freedom from target lesion revascularization (fTLR), occurrence of device- or procedure-related serious adverse events (SAEs), and stent fractures. : A total of 321 patients were enrolled and were a mean age of 73.9 ± 8.7 years; 77.3% were men and 26.5% had chronic limb-threatening ischemia (CLTI). The mean lesion length was 23.6 ± 6.6 cm and the frequency with TASC II C/D lesions and chronic total occlusions was 86.6% and 70.4%, respectively. The Kaplan-Meier estimated PP, PAP, SP, and fTLR at 5 years was 62.4%, 74.1%, 82.3%, and 75.9%, respectively. The mean ankle-brachial index was 0.92 ± 0.15 and the mean improvement in Rutherford class was 2.3 ± 1.4, which was maintained through 5 years. The rate of cumulative device- or procedure-related SAEs through 5 years was 19.9% with only 9.3% of those occurring after the first year. No stent fractures were observed through 5 years by x-ray evaluation. : The 5-year safety and efficacy outcomes of the endoprosthesis were clinically acceptable for treating complex FP lesions in a real-world cohort of Japanese patients with PAD. (.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11323428PMC
http://dx.doi.org/10.1177/1358863X241233528DOI Listing

Publication Analysis

Top Keywords

gore viabahn
8
viabahn endoprosthesis
8
endoprosthesis treatment
8
complex femoropopliteal
8
femoropopliteal lesions
8
postmarket surveillance
8
surveillance study
8
symptomatic patients
8
patients pad
8
device- procedure-related
8

Similar Publications

Article Synopsis
  • Endovascular procedures like TAVR, TEVAR, and EVAR are becoming more common but still carry risks, especially for older and frail patients.
  • This paper discusses two cases where complications arose from large bore access during such procedures and the innovative methods used to manage them.
  • The insights gained from these cases provide valuable guidance for clinicians facing similar challenges in future interventions.
View Article and Find Full Text PDF
Article Synopsis
  • * Methods: Researchers reviewed data from patients who received VBX stent grafts during F/B-EVAR between January 2019 and May 2023, focusing on technical success, potential complications, and overall patient survival.
  • * Results: The study found a 97.5% technical success rate for VBX stent graft implantation, with low rates of complications and high overall survival at 86.7% after
View Article and Find Full Text PDF

Peripheral arterial disease (PAD) affects more than 230 million adults worldwide. Revascularization via angioplasty is a common method to manage stenosis in the superficial femoral artery (SFA). In-stent restenosis, however, is a common complication in endovascular interventions, especially in the SFA.

View Article and Find Full Text PDF

Carotid blowout syndrome (CBS) is a critical and often fatal complication associated with head and neck malignancies. CBS is characterized by 3 stages as follows: threatened, impending, and acute rupture or hemorrhage. The condition is notably linked with a history of radiotherapy for head and neck cancers, with an incidence of approximately 3-4% among those who have received such treatments.

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to compare the biomechanical properties of two types of balloon-expandable covered stents, VBX and iCover, focusing on their use as bridging stents in fenestrated aortic endoprostheses.
  • Three biomechanical tests were conducted on 12 stents to assess their bending strength, pullout force, and shear-stress force by using a dynamometer and analyzing the data microscopically.
  • Results showed that while VBX required more force to bend when crimped, it needed less force when expanded compared to iCover, with both types showing similar pullout force results but different shear-stress forces at various deformations.
View Article and Find Full Text PDF