Introduction: This study aimed to compare the biomechanical properties of two types of intramedullary nails - reconstruction nails (RCN) and cephalomedullary nails (CMN) - each with different proximal fixations, in a model of an osteoporotic subtrochanteric femoral fracture. This study focused on assessing stiffness and load to failure of RCN and CMN nails to provide insight into their clinical applications in osteoporotic fracture treatments.

Materials And Methods: Ten synthetic osteoporotic femoral models were used to generate a comminuted subtrochanteric fracture model. Five femurs were fixed using an RCN, and the remaining five were fixed using a CMN. The constructs were subjected to axial compression to measure their structural stiffness, load to failure, and failure modes.

Results: The CMN group demonstrated a slightly higher load to failure (mean, 2250 N) than the RCN group (mean, 2100 N), which was statistically significant (p = 0.008). However, the stiffness in both groups was statistically similar (RCN, 250 N/mm; CMN, 255 N/mm; p = 0.69). Both groups showed a load to failure exceeding 1500 N, a typically exerted load on the femoral head by a 75 kg individual. The failure patterns differed, with CMN failures starting at the nail insertion area and RCN failures starting at the reconstruction screw area.

Conclusion: The RCN offers stiffness comparable to that of the CMN; although its load to failure is slightly lower than that of the CMN, it still exceeds the physiological tolerance limit. These findings suggest that the RCN is a viable alternative for treating osteoporotic subtrochanteric fractures.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2024.111512DOI Listing

Publication Analysis

Top Keywords

load failure
20
osteoporotic subtrochanteric
12
cephalomedullary nails
8
subtrochanteric fractures
8
rcn
8
cmn
8
stiffness load
8
failures starting
8
failure
7
load
6

Similar Publications

Objectives: We assessed HIV-1 drug resistance profiles among people living with HIV (PLWH) with detectable viral load (VL) and on dolutegravir-based antiretroviral therapy (ART) in Botswana.

Methods: The study utilised available 100 residual HIV-1 VL samples from unique PLWH in Francistown who had viraemia at-least 6 months after initiating ART in Botswana's national ART program from November 2023 to January 2024. Viraemia was categorized as low-level viraemia (LLV) (VL: 200-999 copies/mL) or virologic failure (VF) (VL ≥1000 copies/mL).

View Article and Find Full Text PDF

Aims: Patients with heart failure (HF) often experience delayed identification of palliative care needs. While communication with HF patients and their caregivers is increasingly stressed, systematic conversations about end-of-life care wishes remain a gap. This study explores a dyad experience of Advance Care Planning (ACP) conversations in an HF outpatient clinic.

View Article and Find Full Text PDF

Background: A lateral extra-articular tenodesis (LET) is increasingly being utilized to augment an anterior cruciate ligament reconstruction because it has been shown to reduce the risk of postreconstruction graft failure or recurrent rotatory instability. Various femoral fixation techniques are available, including the use of an interference screw, staple, or suture anchor.

Purpose: To determine and compare the biomechanical properties of an LET graft when using an interference screw, staple, or suture anchor for the femoral fixation for a modified Lemaire LET.

View Article and Find Full Text PDF

Background: The aim of this study was to assess whether there are any differences in plantar pressure distribution and gait cycle between patients with the same implant design and patients with different designs in staged bilateral TKA and to examine the impact of these differences on clinical and functional outcomes.

Methods: A retrospective review was conducted of all patients who underwent staged bilateral TKA. The patients were divided into two groups based on the prosthesis design used in both knees.

View Article and Find Full Text PDF

Collagen/polyvinyl alcohol scaffolds combined with platelet-rich plasma to enhance anterior cruciate ligament repair.

Biomater Adv

December 2024

College of Biological Science and Engineering, Fuzhou University, No. 2 Xueyuan Road, Fuzhou 350108, China; Fujian Key Laboratory of Medical Instrument and Pharmaceutical Technology, Fuzhou University, No. 2 Xueyuan Road, Fuzhou 350108, China. Electronic address:

In anterior cruciate ligament (ACL) repair methods, the continuous enzymatic erosion of synovial fluid can impede healing and potentially lead to repair failure, as well as exacerbate articular cartilage wear, resulting in joint degeneration. Inspired by the blood clot during medial collateral ligament healing, we developed a composite scaffold comprising collagen (1 %, w/v) and polyvinyl alcohol (5 %, w/v) combined with platelet-rich plasma (PRP). The composite scaffold provides a protective barrier against synovial erosion for the ruptured ACL, while simultaneously facilitating tissue repair, thereby enhancing the efficacy of ACL repair techniques.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!