Long-term follow-up of total hip arthroplasty using polyethylene-ceramic composite (sandwich) liner.

Hip Int

Clinical Research Support Unit (HUB-IDIBELL: Bellvitge University Hospital & Bellvitge Biomedical Research Institute), Clinical Pharmacology Department, Bellvitge University Hospital, L'Hospitalet de Llobregat, Barcelona, Spain.

Published: July 2024

Introduction: Ceramic-on-ceramic bearings have been widely used since their introduction in the 1970s. First-generation ceramics have very high fracture (breakage) rates. To overcome this, in the 1990s, modular cotyloid insert designs were developed, consisting of a ceramic and polyethylene composite (sandwich-type) liner; however, high implant fracture rates were observed in the medium term. We aimed to estimate the cumulative incidences of revision surgery (implant failure) and implant failure due to fractures, survival rates (time-to-revision surgery and time-to-fracture), and the long-term clinical and radiological outcomes in our series.

Methods: This was an observational, longitudinal, ambispective, single-centre study based on patients who underwent primary total hip arthroplasty (THA) using a sandwich-type liner (Cerasul), only available in our tertiary hospital between January 1999 and December 2002. Cumulative incidences were estimated and the 95% confidence interval (95% CI) was calculated. The Kaplan-Meier method was used to assess the time-to-revision surgery and time-to-fracture.

Results: 54 patients (49 men) were included, accounting for 59 sandwich-type linear implants. The mean (range) age was 47.4 (22-57) years. The primary THA indications were osteoarthritis (28 patients), osteonecrosis (14), childhood pathology sequelae (11), and inflammatory arthritis (6). The cumulative incidence of revision surgery by implants was 8.5% (5/59, 95% CI, 3.5-19.2%), 9.3% by patients (5/54, 95% CI, 4.0-19.9%), and 5.1% by implant fractures (3/59, 95%CI, 1.7-13.9%). The median (Interquartile Range, IQR) time-to-revision surgery was 158 (72.5-161) months, and the time to fracture was 182 (138-215) months. All primary THAs had good clinical and long-term survival outcomes. All implants had signs of solid fixation.

Conclusions: After a 20-year follow-up period, the polyethylene-ceramic sandwich-type liner showed a long survival rate and low cumulative incidence of implant fracture; however, implant fractures remain the main complication. Orthopaedic surgeons should be aware that some patients still have this type of prosthesis and must be capable of responding quickly if a fracture occurs.

Download full-text PDF

Source
http://dx.doi.org/10.1177/11207000241239624DOI Listing

Publication Analysis

Top Keywords

sandwich-type liner
12
time-to-revision surgery
12
total hip
8
hip arthroplasty
8
implant fracture
8
cumulative incidences
8
revision surgery
8
implant failure
8
cumulative incidence
8
implant fractures
8

Similar Publications

Long-term follow-up of total hip arthroplasty using polyethylene-ceramic composite (sandwich) liner.

Hip Int

July 2024

Clinical Research Support Unit (HUB-IDIBELL: Bellvitge University Hospital & Bellvitge Biomedical Research Institute), Clinical Pharmacology Department, Bellvitge University Hospital, L'Hospitalet de Llobregat, Barcelona, Spain.

Introduction: Ceramic-on-ceramic bearings have been widely used since their introduction in the 1970s. First-generation ceramics have very high fracture (breakage) rates. To overcome this, in the 1990s, modular cotyloid insert designs were developed, consisting of a ceramic and polyethylene composite (sandwich-type) liner; however, high implant fracture rates were observed in the medium term.

View Article and Find Full Text PDF

Facile fabrication of an ultrasensitive sandwich-type electrochemical immunosensor for the quantitative detection of alpha fetoprotein using multifunctional mesoporous silica as platform and label for signal amplification.

Talanta

November 2014

Key Laboratory of Chemical Sensing & Analysis in Universities of Shandong, School of Chemistry and Chemical Engineering, University of Jinan, Jinan 250022, PR China. Electronic address:

A novel and ultrasensitive sandwich-type electrochemical immunosensor was designed for the quantitative detection of alpha fetoprotein (AFP) using multifunctional mesoporous silica (MCM-41) as platform and label for signal amplification. MCM-41 has high specific surface area, high pore volume, large density of surface silanol groups (SiOH) and good biocompatibility. MCM-41 functionalized with 3-aminopropyltriethoxysilane (APTES), gold nanoparticles (Au NPs) and toluidine blue (TB) could enhance electrochemical signals.

View Article and Find Full Text PDF

Ultrasensitive sandwich-type electrochemical immunosensor based on a novel signal amplification strategy using highly loaded toluidine blue/gold nanoparticles decorated KIT-6/carboxymethyl chitosan/ionic liquids as signal labels.

Biosens Bioelectron

November 2014

Key Laboratory of Chemical Sensing & Analysis in Universities of Shandong, School of Chemistry and Chemical Engineering, University of Jinan, Jinan 250022, PR China. Electronic address:

An ultrasensitive sandwich-type electrochemical immunosensor for the quantitative detection of carcinoembryonic antigen (CEA) was designed based on a novel signal amplification strategy. Gold nanoparticles decorated mesoporous silica KIT-6 (Au@KIT-6) with large specific surface area and good adsorption properties was used as a label matrix to immobilize both the secondary antibodies (Ab2) and the electron transfer mediator toluidine blue (TB). Ab2 was loaded on KIT-6 due to the presence of gold nanoparticles (Au NPs) and TB was loaded on KIT-6 by physical adsorption.

View Article and Find Full Text PDF

Due to its advantages, ceramic-on-ceramic bearings have been widely used in young patients for almost 30 years. Long-term survivorship, low wear, and low biological reactivity to particles are some of its characteristics. Even though this material has had a lot of improvements, the risk of fracture is one of the concerns.

View Article and Find Full Text PDF

The mechanisms leading to systematic failure in modular acetabular components with a sandwich insertion (alumina/polyethylene/titanium) have been reconsidered in light of the newly collected Raman spectroscopic results. Raman assessments were conducted on the polyethylene shells, which belonged to a series of six failed sandwich implants with in vivo lifetimes ranging between 2 and 9yr. With only one exception, all implants commonly showed dislodgment of the polyethylene shell during radiographic analyses prior to revision surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!