  • The study aimed to find the best conditions for coronary computed tomography angiography (CCTA) using high-definition (HD) mode and deep learning image reconstruction (DLIR).
  • Experiments were conducted using a 256-row detector CT scanner with various imaging parameters, measuring tasks like image quality, noise levels, and detectability.
  • Results indicated that HD mode combined with DLIR was superior for in-plane image quality, while NR mode with DLIR excelled in noisier environments.

Article Abstract

Purpose: To verify the optimal imaging conditions for coronary computed tomography angiography (CCTA) examinations when using high-definition (HD) mode and deep learning image reconstruction (DLIR) in combination.

Method: A chest phantom and an in-house phantom using 3D printer were scanned with a 256-row detector CT scanner. The scan parameters were as follows - acquisition mode: ON (HD mode) and OFF (normal resolution [NR] mode), rotation time: 0.28 s/rotation, beam coverage width: 160 mm, and the radiation dose was adjusted based on CT-AEC. Image reconstruction was performed using ASiR-V (Hybrid-IR), TrueFidelity Image (DLIR), and HD-Standard (HD mode) and Standard (NR mode) reconstruction kernels. The task-based transfer function (TTF) and noise power spectrum (NPS) were measured for image evaluation, and the detectability index (d') was calculated. Visual evaluation was also performed on an in-house coronary phantom.

Result: The in-plane TTF was better for the HD mode than for the NR mode, while the z-axis TTF was lower for DLIR than for Hybrid-IR. The NPS values in the high-frequency region were higher for the HD mode compared to those for the NR mode, and the NPS was lower for DLIR than for Hybrid-IR. The combination of HD mode and DLIR showed the best value for in-plane d', whereas the combination of NR mode and DLIR showed the best value for z-axis d'. In the visual evaluation, the combination of NR mode and DLIR showed the best values from a noise index of 45 HU.

Conclusion: The optimal combination of HD mode and DLIR depends on the image noise level, and the combination of NR mode and DLIR was the best imaging condition under noisy conditions.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.2024-1353DOI Listing

Publication Analysis

Top Keywords

combination mode
20
mode dlir
20
mode
16
dlir best
16
image reconstruction
12
dlir
9
optimal imaging
8
imaging conditions
8
conditions coronary
8
coronary computed
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!