Purpose: Refixation of acute anterior cruciate ligament (ACL) tears represents an increasingly popular treatment option. Systematic evaluations of various suture technique parameters are still pending. We therefore aimed to evaluate the mechanical pull-out outcomes of various suture methods for optimization of ACL refixation.
Methods: Sixty fresh knees from mature domestic pigs were dissected and the femoral attachment of the ACL was peeled off. The 60 knees were divided in 10 groups and sutured as follows: (A) one suture (1, 2, 4 and 6 passes), (B) two sutures (2, 4 and 6 passes each; sutures knotted together as a loop) and (C) two sutures (2, 4 and 6 passes each, sutures knotted separately). The pull-out test was conducted using a validated electrodynamic testing machine. First occurrence of failure, maximum pull-out load and stiffness were measured. Suture failure was defined as pull-out of the ACL.
Results: Two-point fixation, using two sutures, with at least two passes, showed the most favourable biomechanical stability. The maximum pull-out load was significantly higher with two sutures (529.5 N) used compared to one (310.4 N), < 0.001. No significant differences were found for maximum pull-out loads between two-point fixation versus one-point fixation but stiffness was significantly higher with two-point fixation (107.4 N/mm vs. 79.4 N/mm, < 0.001). More passes resulted in higher maximum pull-out loads.
Conclusion: The results suggest using two independent sutures, refixed separately and at least two suture passes, is appropriate for ACL refixation. More suture passes provide additional strength but are technically challenging to achieve during surgery.
Level Of Evidence: Level IV.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10941523 | PMC |
http://dx.doi.org/10.1002/jeo2.12011 | DOI Listing |
Eur J Ophthalmol
January 2025
Department of Ophthalmology, King's College Hospital, London, UK.
A 42 year old Afro-Caribbean man underwent Baerveldt Glaucoma Implant (BGI) surgery for silicone oil induced glaucoma. Three months following initial surgery, the 3-0 prolene ripcord suture was removed. Anterior segment OCT demonstrates the position of the intracameral portion of the tube before and after the 3/0 prolene stent suture (PSS) removal.
View Article and Find Full Text PDFArthrosc Tech
December 2024
Department of Orthopaedics, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
Tearing of the subscapularis tendon is a common shoulder injury that typically requires arthroscopic repair. The suture-passing device is a standard tool for repairing the subscapularis tendon. However, it poses the risk of device breakage and may cause additional damage to the tendon.
View Article and Find Full Text PDFJBJS Essent Surg Tech
January 2025
The Ohio State University College of Medicine, Columbus, Ohio.
Background: An all-inside endoscopic flexor hallucis longus (FHL) tendon transfer is indicated for the treatment of chronic, full-thickness Achilles tendon defects. The aim of this procedure is to restore function of the gastrocnemius-soleus complex while avoiding the wound complications associated with open procedures.
Description: This procedure can be performed through 2 endoscopic portals, a posteromedial portal (the working portal) and a posterolateral portal (the visualization portal).
Asian J Endosc Surg
December 2024
Division of Gastrointestinal and Pediatric Surgery, Department of Surgery, School of Medicine, Faculty of Medicine, Tottori University, Yonago, Japan.
Internal hernia following colorectal surgery is an uncommon but serious complication. Most reported cases have involved hernias resulting from mesenteric defects after left-sided colon resection. We herein report a case of laparoscopic repair of an internal hernia at the pelvic floor following low anterior resection.
View Article and Find Full Text PDFIn surgery for acute type A aortic dissection, controlling bleeding from the posterior wall of the proximal anastomosis is particularly challenging. To address this, we use the "reversed turn-up technique." For the reinforcement of the proximal aortic stump, Teflon felt strips were placed inside and outside the suture line with 4-0 polypropylene continuous transverse mattress sutures, and BioGlue was applied to the false lumen.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!