Clinical Relevance: Multidiagnostic systems have recently appeared on the market. Knowledge of the repeatability and validity of any instrument is mandatory before its introduction in clinical practice.

Background: The aim of this work is to examine the intrasession repeatability of anterior pole measurements provided by the multidiagnostic device Wave Analyzer Medica 700 (WAM700) and agreement with Pentacam measurements in normal eyes.

Methods: In the right eyes of 113 participants, three repeat measurements of central keratometry, central corneal thickness, anterior chamber depth and corneal eccentricity were made with the WAM700 and Pentacam in random order. Intrasession repeatability and agreement were determined.

Results: Employing WAM700, intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth was good (ICCs ≥ 0.992; CV 0.48-0.98%), yet worse than the values obtained for the Pentacam (ICCs ≥ 0.998; CV 0-0.33%). WAM700 showed excellent intrasession repeatability when used to measure the anterior chamber depth (Sw 0.03 mm). However, the repeatability of this device was inferior for central corneal thickness (Sw 4.24 μm) and keratometry measurements (Sw < 0.21 D) and was poor for corneal eccentricity (Sw 0.07; ICC 0.908; CV 14.58%). Agreement between WAM700 and Pentacam showed a high ICC for the keratometry measurements, central corneal thickness and anterior chamber depth (>0.972) but lower for corneal eccentricity (ICC 0.762).

Conclusions: In healthy eyes, the WAM700 multidiagnostic device showed good intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth measurements. Agreement between WAM700 and Pentacam was good for the anterior chamber depth measurement. However, these instruments cannot be considered interchangeable for keratometry, central corneal thickness and eccentricity readings.

Download full-text PDF

Source
http://dx.doi.org/10.1080/08164622.2024.2324978DOI Listing

Publication Analysis

Top Keywords

intrasession repeatability
24
central corneal
20
corneal thickness
20
anterior chamber
20
chamber depth
20
keratometry central
16
multidiagnostic device
12
thickness anterior
12
repeatability agreement
8
corneal
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!