  • A study was conducted on 40 cardiac resynchronization therapy (CRT) patients to optimize atrial-ventricular delay (AVD) during both atrial sensing (SAVD) and pacing (PAVD) to enhance cardiac function.
  • Researchers used a new electrocardiographic method to measure electrical synchrony and found that optimal PAVD settings were significantly greater than SAVD settings by about 61 ms.
  • The results indicated that when using biventricular pacing, shorter AVDs resulted in improved electrical synchrony, and programming during atrial sensing at longer AVDs led to decreased synchrony.

Article Abstract

Background: Optimization of atrial-ventricular delay (AVD) during atrial sensing (SAVD) and pacing (PAVD) provides the most effective cardiac resynchronization therapy (CRT). We demonstrate a novel electrocardiographic methodology for quantifying electrical synchrony and optimizing SAVD/PAVD.

Methods: We studied 40 CRT patients with LV activation delay. Atrial-sensed to RV-sensed (As-RVs) and atrial-paced to RV-sensed (Ap-RVs) intervals were measured from intracardiac electrograms (IEGM). LV-only pacing was performed over a range of SAVD/PAVD settings. Electrical dyssynchrony (cardiac resynchronization index; CRI) was measured at each setting using a multilead ECG system placed over the anterior and posterior torso. Biventricular pacing, which included multiple interventricular delays, was also conducted in a subset of 10 patients.

Results: When paced LV-only, peak CRI was similar (93 ± 5% vs. 92 ± 5%) during atrial sensing or pacing but optimal PAVD was 61 ± 31 ms greater than optimal SAVD. The difference between As-RVs and Ap-RVs intervals on IEGMs (62 ± 31 ms) was nearly identical. The slope of the correlation line (0.98) and the correlation coefficient r (0.99) comparing the 2 methods of assessing SAVD-PAVD offset were nearly 1 and the y-intercept (0.63 ms) was near 0. During simultaneous biventricular (BiV) pacing at short AVD, SAVD and PAVD programming did not affect CRI, but CRI was significantly (p < .05) lower during atrial sensing at long AVD.

Conclusions: A novel methodology for measuring electrical dyssynchrony was used to determine electrically optimal SAVD/PAVD during LV-only pacing. When BiV pacing, shorter AVDs produce better electrical synchrony.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.14963DOI Listing

Publication Analysis

Top Keywords

cardiac resynchronization
12
atrial-ventricular delay
8
resynchronization therapy
8
atrial sensing
8
ap-rvs intervals
8
pacing
5
determination sensed
4
sensed paced
4
paced atrial-ventricular
4
delay cardiac
4

Similar Publications

Background: Non-response to cardiac resynchronization therapy (CRT) is an important issue in the treatment of heart failure with reduced ejection fraction (HFrEF) and non-left bundle branch block (LBBB). Electrocardiogram-gated myocardial perfusion single-photon emission computed tomography imaging (G-MPI SPECT) is typically used to assess left ventricular (LV) dyssynchrony. This study aimed to determine whether G-MPI parameters are associated with non-responsiveness to CRT.

View Article and Find Full Text PDF

Cardiac resynchronization therapy (CRT) offers significant benefits in symptom alleviation, reduction of rehospitalization rates, and overall survival of patients with heart failure (HF) with reduced ejection fraction (rEF). However, despite its proven efficacy, precisely identifying suitable CRT candidates remains a challenge, with a notable proportion of patients experiencing non-response. Accordingly, many attempts have been made to enhance patient selection, and to identify the best imaging parameters to predict the response and survival after CRT implantation.

View Article and Find Full Text PDF

Left bundle branch block - innocent bystander, silent menace, or both.

Heart Rhythm

December 2024

Christian-Albrechts-University, Medical Faculty, Christian-Albrechts-Platz 4, 24118 Kiel, Germany; University of Applied Science, Life Sciences, An der Karlstadt 8, 27568 Bremerhaven, Germany. Electronic address:

Left bundle branch block (LBBB) causes immediate electrical and mechanical dys-synchrony of the left ventricle (LV) and gradual structural damages in the Purkinje cells and myocardium. Mechanical dys-synchrony reduces the LV ejection fraction (EF) instantly, but only to ≈55% in an otherwise normal heart. Because of the heart's in-built functional redundancy, a patient with LBBB does not always notice the heart's reduced efficiency straight away.

View Article and Find Full Text PDF

Existing techniques for pacing the right ventricle and providing cardiac resynchronization therapy through biventricular pacing are not effective in restoring damage to the conduction system. Therefore, the need for new pacing modalities and techniques with more sensible designs and algorithms is justified. Although the benefits of conduction system pacing (CSP), which mainly include His bundle pacing (HBP) and left bundle branch area pacing (LBBAP), are evident in patients who require conduction system recuperation, the critical criteria for left CSP remain unclear, and the roles of different pacing modalities of CSP for cardiac resynchronization are not definite.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!