Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Significance: Endotracheal intubation is a common approach for airway management in critically ill patients. However, the position of the endotracheal tube (ETT) may be altered during the procedure due to head movements. Accidental displacement or dislodge of the ETT may reduce the airflow, leading to moderate to severe complications, and in some cases even fatality. Therefore, timely detection of changes in ETT position in the trachea is critical to ensure immediate and intermediate interventions to maintain the ETT in the proper position. Currently, there are no widely utilized tools for real-time monitoring of ETT positions.
Aim: The goal of this study is to develop a cost-effective and easy-to-use near-infrared (NIR) device, named Opt-ETT, capable of continuously monitoring the ETT position in the trachea of a patient.
Approach: A side-firing fiber is attached to the side of the ETT to illuminate the trachea tissue with NIR light, and a detector board containing five phototransistors is affixed to the chest skin to measure the intensity of diffusely transmitted light. Displacement of the ETT is estimated using second-order polynomial fitting to the ratios of the phototransistor readings. Monte Carlo simulations, experiment on porcine tissue, and experiments using a swine model have been conducted to assess the feasibility of the device.
Results: The design of the Opt-ETT device has been verified by the Monte Carlo simulations and experiment. The estimation of displacement from experiments using the Opt-ETT exhibited a high degree of agreement with that measured by a reference sensor, with a discrepancy between to within a displacement range from to .
Conclusions: The Opt-ETT device provides a potentially cost-effective solution for real-time and continuous monitoring of ETT position in patient during an intubation procedure.