  • High-precision temperature control of large-area blackbodies is crucial for accurate temperature calibration and thermal imaging, requiring correction of temperature differences between the radiating surface and temperature sensor back surface during testing.* -
  • The article presents a new automated calibration method that uses calibrated infrared thermometers and a three-axis movement system to accurately measure temperatures across different channels of a large-area blackbody.* -
  • Experimental results showed significant improvements: an 85.4% reduction in consistency error, 40.4% better temperature uniformity, and a 43.8% decrease in average measurement deviation, while also reducing calibration time by nearly 10 times under varying conditions.*

Article Abstract

High-precision temperature control of large-area blackbodies has a pivotal role in temperature calibration and thermal imaging correction. Meanwhile, it is necessary to correct the temperature difference between the radiating (surface of use) and back surfaces (where the temperature sensor is installed) of the blackbody during the testing phase. Moreover, large-area blackbodies are usually composed of multiple temperature control channels, and manual correction in this scenario is error-prone and inefficient. At present, there is no method that can achieve temperature-automated calibration for a large-area blackbody radiation source. Therefore, this article is dedicated to achieving temperature-automated calibration for a large-area blackbody radiation source. First, utilizing two calibrated infrared thermometers, the optimal temperature measurement location was determined using a focusing algorithm. Then, a three-axis movement system was used to obtain the true temperature at the same measurement location on a large-area blackbody surface from different channels. This temperature was subtracted from the blackbody's back surface. The temperature difference was calculated employing a weighted algorithm to derive the parameters for calibration. Finally, regarding experimental verification, the consistency error of the temperature measurement point was reduced by 85.4%, the temperature uniformity of the surface source was improved by 40.4%, and the average temperature measurement deviation decreased by 43.8%. In addition, this system demonstrated the characteristics of strong environmental adaptability that was able to perform temperature calibration under the working conditions of a blackbody surface temperature from 100 K to 573 K, which decreased the calibration time by 9.82 times.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10934448PMC
http://dx.doi.org/10.3390/s24051707DOI Listing

Publication Analysis

Top Keywords

large-area blackbody
16
temperature measurement
16
temperature
14
temperature-automated calibration
12
blackbody radiation
12
radiation source
12
temperature control
8
large-area blackbodies
8
temperature calibration
8
temperature difference
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!