  • A study looked at how thickening of the wall between the heart's ventricles (called interventricular septal hypertrophy) might affect patients after a heart procedure called TAVI.
  • Researchers compared 1,033 patients who had TAVI, separating them into two groups: one with normal wall thickness and another with thicker walls.
  • They found that while thicker walls resulted in higher pressure after the procedure, both groups had similar success rates and complications, meaning that wall thickness didn't really change the overall outcome.

Article Abstract

Background: Previous reports suggest septal hypertrophy with an interventricular septum depth (IVSD) ≥ 14 mm may adversely affect outcomes after transcatheter aortic valve implantation (TAVI) due to suboptimal valve placement, valve migration, or residual increased LVOT pressure gradients.

Aims: This analysis investigates the impact of interventricular septal hypertrophy on acute outcomes after TAVI.

Methods: Between 2009 and 2021, 1033 consecutive patients (55.8% male, 80.5 ± 6.7 years, EuroSCORE II 6.3 ± 6.5%) with documented IVSD underwent TAVI at our center and were included for analysis. Baseline, periprocedural, and 30-day outcome parameters of patients with normal IVSD (< 14 mm; group 1) and increased IVSD (≥ 14 mm; group 2) were compared. Data were retrospectively analyzed according to updated Valve Academic Research Consortium-3 (VARC-3) definitions. Comparison of outcome parameters was adjusted for baseline differences between groups using logistic and linear regression analyses.

Results: Of 1033 patients, 585 and 448 patients were allocated to groups 1 and 2, respectively. There was no significant difference between groups regarding transfemoral access rate (82.6% (n = 478) vs. 86.0% (n = 381), p = 0.157). Postprocedural mean transvalvular pressure gradient was significantly increased in group 2 (group 1, 7.8 ± 4.1 mmHg, vs. group 2, 8.9 ± 4.9 mmHg, p = 0.046). Despite this finding, there was no significant difference between groups regarding the rates of VARC-3 adjudicated composite endpoint device success (90.0% (n = 522) vs. 87.6% (n = 388), p = 0.538) or technical success (92.6% (n = 542) vs. 92.6% (n = 415), p = 0.639). Moreover, the groups showed no significant differences regarding the rates of paravalvular leakage ≥ moderate (3.1% (n = 14) vs. 2.6% (n = 9), p = 0.993), postprocedural permanent pacemaker implantation (13.4% (n = 77) vs. 13.8% (n = 61), p = 0.778), or 30-day mortality (5.1% (n = 30) vs. 4.5% (n = 20), p = 0.758).

Conclusion: Although transvalvular mean pressure gradients were significantly higher in patients with increased IVSD after TAVI, acute outcomes were comparable between groups suggesting no early impact of adverse hemodynamics due to elevated IVSD. However, how these differences in hemodynamic findings may affect mid- and long-term outcomes, especially in terms of valve durability, needs to be evaluated in further investigations.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00392-024-02432-3DOI Listing

Publication Analysis

Top Keywords

septal hypertrophy
12
transcatheter aortic
8
aortic valve
8
valve implantation
8
valve
4
implantation patients
4
patients septal
4
hypertrophy background
4
background previous
4
previous reports
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!