Peripheral defocus of monofocal intraocular lenses.

J Cataract Refract Surg

From the Ophthalmic Biophysics Laboratory, L V Prasad Eye Institute, Hyderabad, Telangana, India (Natarajan, Singh, Mohamed, Vadavalli); Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami, Miller School of Medicine, Miami, Florida (Heilman, Ruggeri, Parel, Manns); Department of Biomedical Engineering, University of Miami College of Engineering, Coral Gables, Florida (Heilman, Ruggeri, Parel, Manns); Brien Holden Vision Institute Limited, Sydney, New South Wales, Australia; School of Optometry and Vision Science, The University of New South Wales, Sydney, New South Wales, Australia (Ho); Cataract & Refractive surgery services, L V Prasad Eye Institute, Hyderabad, Telangana, India (Singh, Reddy, Vadavalli); The Shantilal Shanghvi Cornea Institute, L V Prasad Eye Institute, Hyderabad, Telangana, India (Vadavalli).

Published: June 2024

Purpose: To quantify the angular dependence of monofocal intraocular lens (IOL) power.

Setting: Ophthalmic Biophysics Laboratory, Kallam Anji Reddy campus, L V Prasad Eye Institute, Hyderabad, India.

Design: Laboratory study.

Methods: Experiments were performed on IOLs from 2 different manufacturers (APPALENS 207, Appasamy Associates and SN60WF, Alcon Laboratories, Inc.). IOL powers ranged from 17 to 25 diopters (D). The IOLs were mounted in a fluid-filled chamber, and the on-axis and off-axis powers were measured using a laser ray tracing system over the central 3 mm zone with delivery angles ranging from -30 to +30 degrees in 5-degree increments. The position of the best focus was calculated for each IOL at each angle. The angular dependence of IOL power was compared with theoretical predictions.

Results: Peripheral defocus increased significantly with increasing incidence angle and power. The peripheral defocus at ±30 degrees increased from 5.8 to 8.5 D when the power increased from 17.5 to 24.5 D for APPALENS 207 and from 4.9 to 7.4 D when the power increased from 17 to 25 D for SN60WF. The mean difference between the measured and theoretical tangential power at ±30 degrees was 0.50 ± 0.16 D for the APPALENS 207 and -0.40 ± 0.10 D for the SN60WF, independent of IOL power.

Conclusions: IOLs introduce a significant amount of peripheral defocus which varies significantly with IOL power and design. Given that peripheral defocus is related to lens power, replacement of the crystalline lens (approximately 24 D) with an IOL will produce a significant difference in peripheral defocus profile after surgery.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11146172PMC
http://dx.doi.org/10.1097/j.jcrs.0000000000001441DOI Listing

Publication Analysis

Top Keywords

peripheral defocus
24
appalens 207
12
monofocal intraocular
8
angular dependence
8
lens iol
8
iol power
8
±30 degrees
8
power increased
8
iol
7
power
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!