Introduction: Transsubclavian venous implantation of the Aveir leadless cardiac pacemaker (LCP) has not been previously reported.

Methods And Results: Three cases of transsubclavian implantation of the Aveir LCP are reported. Two cases were postbilateral orthotopic lung transplant, without appropriate femoral or jugular access due to recent ECMO cannulation and jugular central venous catheters. In one case, there was strong patient preference for same-day discharge. Stability testing confirmed adequate fixation and electrical testing confirmed stable parameters in all cases. All patients tolerated the procedure well without significant immediate complications.

Conclusions: We demonstrate the feasibility of transsubclavian implantation of the Aveir LCP.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.16241DOI Listing

Publication Analysis

Top Keywords

implantation aveir
16
transsubclavian venous
8
venous implantation
8
aveir leadless
8
leadless cardiac
8
transsubclavian implantation
8
aveir lcp
8
testing confirmed
8
first-in-man report
4
transsubclavian
4

Similar Publications

Leadless pacemakers (LPs) are emerging options for bradyarrhythmias. However, extraction can be risky if the device is in an unfavorable position. We present a challenging case of a Nanostim LP (NLP) (Abbott Medical Inc.

View Article and Find Full Text PDF

3D computed tomography integration guiding permanent Aveir AR leadless pacemaker implantation: a case report.

Eur Heart J Case Rep

January 2025

AL Qassimi Hospital, Emirates Health Services (EHS), Wasit Street-Al khazamiya, Sharjah 3500, United Arab Emirates.

Background: The use of single-chamber, right ventricular (RV) leadless pacemakers (LPs) has been well established, the introduction of a right atrial LPs has opened the door for dual-chamber leadless pacing. Cardiac computed tomography (CT) segmentation integration might provide proper visual guide during the procedure.

Case Summary: A 58-year-old male patient was brought to the emergency department with dizziness and complete heart block.

View Article and Find Full Text PDF

We present the case of a 64-year-old man who, during the implantation of an active-fixation leadless pacemaker (LP, Aveir VR, Abbott, USA), underwent several external defibrillation shocks up to 240 Joules, due to symptomatic sustained supraventricular tachycardia at 160 bpm. The shocks, delivered both before and after the screwing of the device in the low interventricular septum, did not cause any technical damage to the device, and no complications were observed. The device was then deployed successfully.

View Article and Find Full Text PDF

Minimizing for Maximum Benefit: An Illustrative Case-Series of Atrial Only Leadless Pacing.

Pacing Clin Electrophysiol

January 2025

Division of Cardiology, Department of Medicine, Weill Cornell Medicine, New York Presbyterian Hospital, New York, USA.

Leadless pacing technology now includesdedicated atrial helix-fixation leadless pacemakers (LPs), expanding theapplication of leadless devices for patients with sinus node dysfunction andatrioventricular block during sinus rhythm. This first reportedcase-series of atrial LPs describes and discusses the potential use-casescenarios of recently approved helix-fixation atrial LPs. The article highlights important concepts regarding their use, including implantationtechniques, programming, battery conservation, and the low rate of progressionof AV block in patients implanted with AAI(R) pacemakers.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!