Introduction: Transsubclavian venous implantation of the Aveir leadless cardiac pacemaker (LCP) has not been previously reported.
Methods And Results: Three cases of transsubclavian implantation of the Aveir LCP are reported. Two cases were postbilateral orthotopic lung transplant, without appropriate femoral or jugular access due to recent ECMO cannulation and jugular central venous catheters. In one case, there was strong patient preference for same-day discharge. Stability testing confirmed adequate fixation and electrical testing confirmed stable parameters in all cases. All patients tolerated the procedure well without significant immediate complications.
Conclusions: We demonstrate the feasibility of transsubclavian implantation of the Aveir LCP.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/jce.16241 | DOI Listing |
Zhonghua Er Ke Za Zhi
January 2025
Department of Cardiology, Shanghai Children's Medical Center Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai200127, China.
J Innov Card Rhythm Manag
December 2024
Department of Electrophysiology, MedStar Health: Heart and Vascular Institute at MedStar Washington Hospital Center, Washington, DC, USA.
Leadless pacemakers (LPs) are emerging options for bradyarrhythmias. However, extraction can be risky if the device is in an unfavorable position. We present a challenging case of a Nanostim LP (NLP) (Abbott Medical Inc.
View Article and Find Full Text PDFEur Heart J Case Rep
January 2025
AL Qassimi Hospital, Emirates Health Services (EHS), Wasit Street-Al khazamiya, Sharjah 3500, United Arab Emirates.
Background: The use of single-chamber, right ventricular (RV) leadless pacemakers (LPs) has been well established, the introduction of a right atrial LPs has opened the door for dual-chamber leadless pacing. Cardiac computed tomography (CT) segmentation integration might provide proper visual guide during the procedure.
Case Summary: A 58-year-old male patient was brought to the emergency department with dizziness and complete heart block.
Pacing Clin Electrophysiol
January 2025
Electrophysiology and Cardiac Pacing Unit, San Giuliano Hospital, Giugliano in Campania, Italy.
We present the case of a 64-year-old man who, during the implantation of an active-fixation leadless pacemaker (LP, Aveir VR, Abbott, USA), underwent several external defibrillation shocks up to 240 Joules, due to symptomatic sustained supraventricular tachycardia at 160 bpm. The shocks, delivered both before and after the screwing of the device in the low interventricular septum, did not cause any technical damage to the device, and no complications were observed. The device was then deployed successfully.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2025
Division of Cardiology, Department of Medicine, Weill Cornell Medicine, New York Presbyterian Hospital, New York, USA.
Leadless pacing technology now includesdedicated atrial helix-fixation leadless pacemakers (LPs), expanding theapplication of leadless devices for patients with sinus node dysfunction andatrioventricular block during sinus rhythm. This first reportedcase-series of atrial LPs describes and discusses the potential use-casescenarios of recently approved helix-fixation atrial LPs. The article highlights important concepts regarding their use, including implantationtechniques, programming, battery conservation, and the low rate of progressionof AV block in patients implanted with AAI(R) pacemakers.