Purpose: To maintain safe intrarenal pelvic pressure (IPP), the combination of flexible ureteroscope (fURS) and traditional ureteral access sheath (T-UAS) should maintain a basic rule that is the ratio of endoscope-sheath diameter (RESD) ≤ 0.75. However, the negative-pressure ureteral access sheath (NP-UAS) may break the rule of negative pressure suction. This study aimed to examine the effect of NP-UAS on IPP and flow rate (FR) with varying RESD.
Methods: In a 3D-printed renal model, flexible ureteroscopy lithotripsy (fURL) was replicated. Six sizes of fURS paired with 12Fr T-UAS and NP-UAS resulted in six distinct RESDs of 0.63, 0.78, 0.87, 0.89, 0.90, and 0.91. While the irrigation pressure (IRP) was set between 100 and 800 cmHO and the sucking pressure (SP) was set between 0 and 800 cmHO, the IPP and FR were measured in each RESD.
Results: NP-UASs can reduce the IPP and increase the FR at the same RESD compared to T-UASs. The IPP decreased with increasing SP with NP-UAS. When RESD ≤ 0.78, T-UAS and NP-UAS can maintain IPP < 40 cmHO in most circumstances. When RESD = 0.87, it is challenging for T-UAS to sustain IPP < 40 cmHO; however, NP-UAS can do so. When RESD ≥ 0.89, it is difficult to maintain an IPP < 40 cmHO even with NP-UAS.
Conclusion: NP-UAS can decrease IPP and increase FR compared with T-UAS. To maintain a safe IPP, it is recommended that RESD < 0.85 when utilizing NP-UAS.