Purpose: When performing single-point dose verification in VMAT, it is necessary to avoid the regions with steep dose gradient. We propose a method to obtain the estimated value ( U) of uncertainty of the absolute dose measurement due to the phantom setup error by using dose gradient calculated from treatment planning system (TPS), for evaluating the appropriate measurement points.

Methods: The dose gradient was calculated from the planned dose values in the vicinity of the isocenter point using TPS. The phantom setup error was estimated. The U was calculated using the proposed formula after estimating the phantom setup error. Then, the dose gradient was calculated from the measured dose values in the vicinity of the isocenter point specified by TPS using the Tough water phantom with ionization chamber (IC), and U was calculated as in U.

Results: The correlation coefficient between U and U was 0.984, which indicates a high correlation. The average of the difference between U and U was -0.24%. We considered that this result was caused by the influence of volume averaging effect of IC.

Conclusion: The U obtained from this proposed method reflects the uncertainty of the absolute dose measurement due to the phantom setup error and is useful for evaluating the appropriate measurement points for absolute dose measurement.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.2024-1371DOI Listing

Publication Analysis

Top Keywords

phantom setup
20
setup error
20
absolute dose
16
dose measurement
16
dose gradient
16
measurement phantom
12
gradient calculated
12
dose
11
treatment planning
8
uncertainty absolute
8

Similar Publications

Eye lens dosimetry: does the direction of rotation (vertical or horizontal) play a role in type testing?

J Radiol Prot

January 2025

Radiation Protection Dosimetry (6.3), Physikalisch-Technische Bundesanstalt, Braunschweig, NDS, GERMANY.

With the International Commission on Radiological Protection (ICRP) lowering the annual dose limit for the eye lens to 20 mSv, precise monitoring of eye lens exposure has become essential. The personal dose equivalent at a depth of 3 mm, Hp(3), is the measurement method for monitoring the dose to the lens of the eye. Traditional dosimetry methods primarily address lateral radiation exposure scenarios, where radiation approaches from the left or right, necessitating the rotation of the phantom during type testing around the vertical axis.

View Article and Find Full Text PDF

Background: Dosimetric commissioning and quality assurance (QA) for linear accelerators (LINACs) present a significant challenge for clinical physicists due to the high measurement workload and stringent precision standards. This challenge is exacerbated for radiosurgery LINACs because of increased measurement uncertainty and more demanding setup accuracy for small-field beams. Optimizing physicists' effort during beam measurements while ensuring the quality of the measured data is crucial for clinical efficiency and patient safety.

View Article and Find Full Text PDF

To accurately model and validate the 6 MV Elekta Compactlinear accelerator using the Geant4 Application for Tomographic Emission (GATE). In particular, this study focuses on the precise calibration and validation of critical parameters, including jaw collimator positioning, electron source nominal energy, flattening filter geometry, and electron source spot size, which are often not provided in technical documentation. Methods: Simulation of the Elekta Compact6 MV linear accelerator was performed using the Geant4 Application for Tomographic Emission (GATE) v.

View Article and Find Full Text PDF

Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.

Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.

View Article and Find Full Text PDF

Set-up errors are a problem for pre-clinical irradiators that lack imaging capabilities. The aim of this study was to investigate the impact of the potential set-up errors on the dose distribution for a mouse with a xenographic tumour irradiated with a standard Cs-137 cell irradiator equipped with an in-house lead collimator with 10 mm diameter apertures. The EGSnrc Monte-Carlo (MC) code was used to simulate the potential errors caused by displacements of the mouse in the irradiation setup.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!