AI Article Synopsis

  • The study focuses on the fixation of the trochlea in low transcondylar fractures of the distal humerus, which is crucial for elbow stability during surgery.
  • Twenty human cadaveric humeri were tested, divided into two groups, with one group utilizing screws that extended into the trochlea and the other not.
  • Results showed that the group with trochlea fixation had significantly less displacement and higher ultimate failure loads, indicating improved stability in the surgical construct.

Article Abstract

Objective: The trochlea is of great importance for the stability of the elbow and its fixation in low transcondylar fractures of the distal humerus is especially challenging. The aim of this study was to determine the optimal trochlea fixation in double plate osteosynthesis of intraarticular distal humerus fractures.

Methods: A low transcondylar, C3-type distal humerus fracture was created in 20 fresh-frozen human cadaveric humeri. The samples were then randomly divided into two groups of 10 specimens each. Double plate osteosynthesis was performed in both groups. In group A, the two most distal screws of the lateral plate were inserted into the trochlea fragment. In group B, these screws did not extend into the trochlea. Displacement under cyclic loading and ultimate failure loads were determined for all specimens.

Results: Group A showed significantly less displacement under cyclic loading in each measurement interval (0.92 mm vs. 1.53 mm after 100 cycles, p = 0 0.006; 1.10 mm vs. 1.84 mm after 1000 cycles, p = 0.007; 1.18 mm vs. 1.98 mm after 2000 cycles, p = 0.008). The ultimate failure load was significantly higher in group A than in group B (345.61 ± 120.389 N vs. 238.42 ± 131.61 N, p = 0.037).

Conclusions: Fixation of the trochlea with interdigitating screws in double plate osteosynthesis of low-condylar type C distal humerus fractures results in superior construct stability.

Level Of Evidence: not applicable (biomechanical).

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2024.111486DOI Listing

Publication Analysis

Top Keywords

distal humerus
20
double plate
16
plate osteosynthesis
16
low transcondylar
12
fixation trochlea
8
humerus fractures
8
displacement cyclic
8
cyclic loading
8
ultimate failure
8
trochlea
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!