Evaluation of silicon and indium gallium arsenide photodiodes as direct timing detectors for pulsed x-ray systems.

Rev Sci Instrum

C-PCS, Physical Chemistry and Applied Spectroscopy, Los Alamos National Laboratory, Los Alamos, New Mexico 87545, USA.

Published: March 2024

Benchtop pulsed x-ray systems are commonly used to record dynamic material data on the order of nanoseconds, but pulse timing is often difficult to accurately determine. This study demonstrates that commercially available photodiodes can be used effectively for direct x-ray pulse detection without the need for visible light scintillators. X-ray pulses from four commercially available flash x-ray systems were quantified using one silicon and two indium gallium arsenide (InGaAs) photodiodes. The measured InGaAs pulse durations were strongly dependent on radiation dose in the non-linear operating regime, so the photodiodes were shielded to operate below the 2.5 V non-linear regime threshold. The average pulse duration and pulse arrival time jitter of the photodiodes for each x-ray source were within several nanoseconds with the exception of two sets of measurements that were affected by low instrument sensitivity and electrical noise. These results show that InGaAs photodiodes can be used as effective and repeatable stand-alone timing diagnostics for x-ray pulses as short as 20 ns or less.

Download full-text PDF

Source
http://dx.doi.org/10.1063/5.0191520DOI Listing

Publication Analysis

Top Keywords

x-ray systems
12
silicon indium
8
indium gallium
8
gallium arsenide
8
pulsed x-ray
8
x-ray pulses
8
ingaas photodiodes
8
x-ray
7
photodiodes
6
pulse
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!