Purpose: The small size of the cochlea, and its location deeply embedded in thick temporal bone, poses a challenge for intra-cochlear guidance and diagnostics. Current radiological imaging techniques are not able to visualize the cochlear microstructures in detail. Rotational optical coherence tomography (OCT) fibers show great potential for intra-cochlear guidance. The generated images could be used to map, and study, the tiny cochlear microstructures relevant for hearing.

Methods: This work describes the design of a rotational OCT probe with an outer diameter of 0.9 mm. It further discusses a robotic system, which features a remote center of motion mechanism, dedicated to the probe's positioning, fine manipulation and stable insertion into the cochlear micro-spaces. Furthermore, the necessary calibration steps for 3D reconstruction are described, followed by a detailed quantitative analysis, comparing the 3D reconstructions using a synthetic, 2:1 scaled scala tympani model with a reconstruction from micro-CT, serving as the ground truth. Finally, the potential of the system is demonstrated by scanning a single ex vivo cadaveric human cochlea.

Results: The study investigates five insertions in the same 2:1 scaled tympani model, along with their corresponding 3D reconstruction. The comparison with micro-CT results in an average root-mean-square error of 74.2 µm, a signed distance error of 38.1 µm and a standard deviation of 63.6 µm. The average F-score of the reconstructions, using a distance threshold of 100 and 74.2 µm, resulted in 83.0% and 71.8%, respectively. Insertion in the cadaveric human cochlea showed the challenges for straight insertion, i.e., navigating the hook region.

Conclusion: Overall, the system shows great potential for intra-cochlear guidance and diagnostics, due to the system's capability for precise and stable insertion into the basal turn in the scala tympani. The system, combined with the calibration procedure, results in detailed and precise 3D reconstructions.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11548-024-03081-7DOI Listing

Publication Analysis

Top Keywords

intra-cochlear guidance
12
guidance diagnostics
8
cochlear microstructures
8
great potential
8
potential intra-cochlear
8
stable insertion
8
scala tympani
8
tympani model
8
cadaveric human
8
742 µm
8

Similar Publications

OCT-based intra-cochlear imaging and 3D reconstruction: ex vivo validation of a robotic platform.

Int J Comput Assist Radiol Surg

May 2024

Department of Neurosciences, KU Leuven, Herestraat 49, 3001, Leuven, Belgium.

Purpose: The small size of the cochlea, and its location deeply embedded in thick temporal bone, poses a challenge for intra-cochlear guidance and diagnostics. Current radiological imaging techniques are not able to visualize the cochlear microstructures in detail. Rotational optical coherence tomography (OCT) fibers show great potential for intra-cochlear guidance.

View Article and Find Full Text PDF

Auditory performance of post-lingually deafened adult cochlear implant recipients using electrode deactivation based on postoperative cone beam CT images.

Eur Arch Otorhinolaryngol

April 2021

Postgraduate Program at the Department of Ophthalmology, Otorhinolaryngology, Head and Neck Surgery, Ribeirão Preto Medical School, University of São Paulo, Av Bandeirantes, 3900, Ribeirão Preto, São Paulo, Brazil.

Purpose: The use of image processing techniques to estimate the position of intra-cochlear electrodes has enabled the creation of personalized maps to meet the individual stimulation needs of cochlear implant (CI) recipients. The aim of this study was to evaluate a novel technique of electrode deactivation based on postoperative cone beam computed tomography (CBCT) images in post-lingually deafened adult CI recipients.

Methods: Based on postoperative CBCT images, the positioning of the electrodes was estimated in relation to the modiolus in 14 ears of 13 post-lingually deafened adult CI recipients.

View Article and Find Full Text PDF

Characterizing the size of the target region for atraumatic opening of the cochlea through the facial recess.

Comput Med Imaging Graph

October 2019

Department of Otolaryngology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany; Cluster of Excellence EXC 1077/1 "Hearing4all", Hannover Medical School, Germany.

Surgical treatment with a cochlear implant (CI) for hearing rehabilitation requires a highly accurate and personalized opening of the inner ear (cochlea) to protect the delicate intra-cochlear fine structures, whose functional integrity needs to be maintained to preserve residual hearing. Spatial orientation within the complex anatomy of the lateral skull base during the procedure is a highly demanding task for the surgeon. In order to reduce risk of facial nerve palsy and loss of residual hearing as well as to establish minimally invasive CI surgery (minCIS), image-guided procedures incorporating surgical assistance systems are under development.

View Article and Find Full Text PDF

Introduction: Visualization of inner ear structures is a valuable strategy for researchers and clinicians working on hearing pathologies. Optical coherence tomography (OCT) is a high-resolution imaging technology which may be used for the visualization of tissues. In this experimental study we aimed to evaluate inner ear anatomy in well-prepared human labyrinthine bones.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!