Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: To evaluate the intraobserver repeatability and interobserver reproducibility of a newly developed dynamic real-time visualization 25 kHz swept-source optical coherence tomography (SS-OCT) based biometer (ZW-30, TowardPi Medical Technology Ltd, China) and compare its agreement with another SS-OCT based biometer (IOLMaster 700, Carl Zeiss Meditec AG, Jena, Germany).
Methods: Eighty-two healthy right eyes were enrolled in this prospective observational study. Measurements were repeated for three times using the ZW-30 and IOLMaster 700 in a random order. Obtained parameters included axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), astigmatism magnitude (AST), vector J, vector J, and corneal diameter (CD). The within-subject standard deviation (Sw), test-retest (TRT) variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were adopted to assess the intraobserver repeatability and interobserver reproducibility. The double-angle plot was also used to display the distribution of AST. To estimate agreement, Bland-Altman plots were used.
Results: For the intraobserver repeatability and interobserver reproducibility, the Sw, TRT and CoV for all parameters were low. Meanwhile, the ICC values were all close to 1.000, except for the J (ICC = 0.887 for the intraobserver repeatability). The double-angle plot showed that the distribution of AST measured by these two devices was similar. For agreement, the Bland-Altman plots showed narrow 95% limits of agreements (LoAs) for AL, CCT, AQD, ACD, LT, Km AST, J, J, and CD (- 0.02 mm to 0.02 mm, - 7.49 μm to 8.08 μm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.04 mm, - 0.07 mm to 0.08 mm, - 0.16 D to 0.30 D, - 0.30 D to 0.29 D, - 0.16 D to 0.16 D, - 0.23 D to 0.13 D, and - 0.39 mm to 0.10 mm, respectively).
Conclusions: The newly dynamic real-time visualization biometer exhibited excellent intraobserver repeatability and interobserver reproducibility. The two devices both based on the SS-OCT principle had similar ocular parameters measurement values and can be interchanged in clinical practice.