Cardiovascular images: pacemaker-lead fracture and excessive coiling in a dog.

J Vet Cardiol

Department of Clinical Science, College of Veterinary Medicine, North Carolina State University, 1060 William Moore Drive, Raleigh, NC 27606, USA.

Published: April 2024

AI Article Synopsis

  • A seven-year-old Labrador Retriever, previously fitted with a transvenous pacemaker for atrial issues, was brought in for emergency treatment due to suspected pacemaker failure after five years.
  • Upon examination, the dog exhibited a very low heart rate and tests revealed no normal or paced heart activity, confirming a malfunction in the pacemaker's lead.
  • After finding a fractured and excessively coiled lead via X-rays, the veterinarians opted to install a new epicardial pacing system, and while the dog had no immediate complications post-surgery, it eventually needed increased cardiac treatment and was euthanized two years later.

Article Abstract

A seven-year-old male castrated Labrador Retriever presented emergently due to concern for pacemaker malfunction five years after successful transvenous pacemaker implantation to treat partial atrial standstill. On presentation, the dog's pulse rate was 30-50 beats per minute. An electrocardiogram showed no spontaneous atrial activity or paced ventricular activity. Pacemaker interrogation revealed an increased impedance of 7557 ohms, indicating a lead malfunction. Thoracic radiographs confirmed the lead was fractured and had excessive coiling. The transvenous pacing system was turned off, left in place, and an epicardial pacing system was implanted the following day. The dog was discharged with no perioperative complications. The dog eventually required escalated medical therapy for progressive cardiac disease and was euthanized two years after implantation of the replacement pacemaker. This manuscript illustrates a complete lead fracture and excessive lead coiling, which has not previously been detailed in veterinary medicine.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvc.2024.02.004DOI Listing

Publication Analysis

Top Keywords

fracture excessive
8
excessive coiling
8
pacing system
8
cardiovascular images
4
images pacemaker-lead
4
pacemaker-lead fracture
4
coiling dog
4
dog seven-year-old
4
seven-year-old male
4
male castrated
4

Similar Publications

Developing hydrogels with high conductivity and toughness a facile strategy is important yet challenging. Herein, we proposed a new strategy to develop conductive hydrogels by growing metal dendrites. Water-soluble Sn ions were soaked into the gel and then converted to Sn dendrites an electrochemical reaction; the excessive Sn ions were finally removed by water dialysis, accompanied by dramatic shrinkage of the gel.

View Article and Find Full Text PDF

Classic congenital adrenal hyperplasia (CAH) due to 21-hydroxylase deficiency results in severe cortisol and aldosterone deficiency, leading to persistent adrenal stimulation and excess production of ACTH and adrenal androgens. This review examines the clinical considerations and challenges of balancing under- and overtreatment with glucocorticoids in adolescent and adult male individuals with CAH. Adolescents face many unique challenges that can hinder adherence, hormonal control, and transition to independence.

View Article and Find Full Text PDF

Adrenal insufficiency (AI) is a serious disorder characterized by the adrenal glucocorticoid deficiency. Regardless of the etiology, AI patients need long-term replacement therapy for glucocorticoids and, in some cases, for mineralocorticoids. The replacement therapy cannot completely mirror the physiological secretion patterns, and therefore, glucocorticoid excess is a common sequela in AI patients.

View Article and Find Full Text PDF

Research into the Impact of Seawater Ionic Composition on the Rheological Characteristics of a Cationic Guar Gum-Based Fracturing Fluid.

ACS Omega

January 2025

School of Petroleum Engineering, Xi 'an Shiyou University, No. 18, East Section of Electronic second Road, Yanta District, Xi'an, Shaanxi 710054, China.

This study examined the effects of salt content and salt type on the properties of the hydroxypropyl guar gum fracturing fluid. In this study, we conducted a thorough analysis of the impact that various ions in seawater have on the performance of fracturing fluids. We identified the cross-linked polymer that performs optimally at a specific concentration of the binding agent.

View Article and Find Full Text PDF

Background: Observational studies have shown that selective serotonin reuptake inhibitors are associated with an increased risk of bone fractures, but the association can be confounded by indication and other sources of systematic bias that can be minimised in randomised controlled trials (RCTs).

Aim: Our aim was to report the rate, site, context, and predictors of fractures after stroke, and whether the fractures modified the effect of fluoxetine on modified Rankin score (mRS) at six months in an individual patient data meta-analysis of 5907 patients enrolled in three RCTs of fluoxetine (20mg for six months) for stroke recovery.

Methods: We classified fractures by treatment allocation, site (and thus likelihood of osteoporosis) and context, then performed multivariable analyses to explore independent predictors of fractures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!