Reliability of two devices for shoulder strength assessment: Wii Fit Balance Board and hand-held dynamometer.

Shoulder Elbow

Exercise and Rehabilitation Sciences Laboratory, School of Physical Therapy, Faculty of Rehabilitation Sciences, Universidad Andres Bello, Santiago, Chile.

Published: March 2024

Background: The aim of this study was to compare the reliability and agreement between two devices - Wii Fit Balance Board (WBB) versus Hand-Held Dynamometer (HHD) to measure isometric strength during the athletic shoulder (ASH) test in healthy amateur rugby players.

Methods: Fifteen males (23.73 ± 2.8 years) completed two testing sessions. Maximal isometric contractions using the dominant arm (D) and non-dominant arm (ND) against a WBB and HHD were assessed at three angles of abduction (180°, 'I'; 135°, 'Y' and 90°, 'T'), in a prone lying position.

Results: The results indicate a very large correlation between the HHD and the WBB. WBB provides acceptable reliability at I-Test D (CV = 9.97%, ICC = 0.88) and HHD in the I-Test D (CV = 8.90%, ICC = 0.94), I-Test ND (CV = 8.60%, ICC = 0.95) in peak strength values. The HHD is most reliable in D ASH I-Y-T (CV = 10.94%) and WBB (CV = 11.05%). In the ND ASH I-Y-T test, the HHD is the most reliable (CV = 12.5%) compared to the WBB (CV = 14.43%).

Conclusions: These results suggest that WBB is a reliable device to assess strength in the ASH test with a very large correlation with the HHD. WBB and HHD are two affordable devices to assess isometric shoulder strength.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10901178PMC
http://dx.doi.org/10.1177/17585732221145558DOI Listing

Publication Analysis

Top Keywords

shoulder strength
8
wii fit
8
fit balance
8
balance board
8
hand-held dynamometer
8
wbb
8
hhd
8
ash test
8
wbb hhd
8
large correlation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!