Purpose: The steady-state pattern electroretinogram (ssPERG) is used to assess retinal ganglion cell function in a variety of research contexts and diagnostic applications. In certain groups of patients or study participants, stable central fixation of the stimulus is not guaranteed. The present study aimed at assessing the effects of misfixation on the ssPERG response to checkerboard reversal stimuli.

Methods: Using two check sizes (0.8° and 15°), we compared ssPERG responses for several amounts of fixation deviation, ranging from 0° to 19° horizontally and from 0° to 14° diagonally. The stimulus area extended to 15° eccentricity, stimulus reversal rate was 15/s.

Results: Up to around 7° eccentricity, there was no sizable effect of fixation deviation under most conditions. Effects were somewhat larger for nasal than for temporal deviation, in particular for small checks. Diagonal deviation was associated with a response to luminance onset/offset at 7.5 Hz (subharmonic of the reversal rate), most prominently when the interior of a large check was fixated.

Conclusion: Generally, moderate inaccuracies of fixation do not have a sizable effect on ssPERG amplitude. However, with large checks, the luminance response has to be considered.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10954955PMC
http://dx.doi.org/10.1007/s10633-024-09967-wDOI Listing

Publication Analysis

Top Keywords

steady-state pattern
8
pattern electroretinogram
8
fixation deviation
8
reversal rate
8
eccentric fixation
4
fixation steady-state
4
electroretinogram purpose
4
purpose steady-state
4
ssperg
4
electroretinogram ssperg
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!