Aim: To report on the performance characteristics of the 5-ring GE Discovery MI PET/CT systems using the AAPM TG-126 report and compare these results to NEMA NU 2-2012 where applicable.

Materials And Methods: TG-126 testing was performed on two GE 5-Rings Discovery MI scanners. Tests performed included spatial resolution, PET/CT image-registration accuracy, sensitivity, count rate performance, accuracy of corrections, image contrast, scatter/attenuation correction, and image uniformity. All acquired data were analyzed using scanner console or free software tools as described by TG-126 and the results were then compared to published NEMA NU 2-2012 values.

Results: Both scanners gave similar resolution results for TG-126 and NEMA NU 2-2012 and were within manufacturer specifications. Image-registration accuracy between PET and CT using our clinical protocol showed excellent results with values ≤1 mm. Sensitivity using TG-126 was 19.43 cps/kBq while for NEMA the value was 20.73 cps/kBq. The peak noise-equivalent counting rate was 2174 kcps at 63.1 kBq/mL and is not comparable to NEMA NU 2-2012 due to differences in phantoms and methods used to measure and calculate this parameter. The accuracy of corrections for count losses for TG-126 were expressed in SUV values and found to be within 10% of the expected SUV measurement of 1. Image contrast and scatter/attenuation correction using the TG-126 method gave acceptable results. Image uniformity assessment resulted in values within the recommended ± 5% limits.

Conclusion: These results show that the 5-ring GE Discovery MI PET/CT scanner testing using TG-126 is reproducible and has similar results to NEMA NU 2-2012 tests where applicable. We hope these results start to form the basis to compare PET/CT systems using TG-126.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11005963PMC
http://dx.doi.org/10.1002/acm2.14315DOI Listing

Publication Analysis

Top Keywords

nema 2-2012
20
5-ring discovery
12
discovery pet/ct
12
tg-126
10
performance characteristics
8
characteristics 5-ring
8
pet/ct scanner
8
aapm tg-126
8
tg-126 report
8
pet/ct systems
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!