Unlabelled: This case report reviews worsening obstructive sleep apnea (OSA) events in a patient over-titrated with a hypoglossal nerve stimulator. A healthy 57-year-old man underwent hypoglossal nerve stimulator implantation for moderate OSA. During an in-laboratory hypoglossal nerve stimulator titration the patient had improvement of his OSA at 2.1 V. However, a further increase of voltage resulted in worsening of obstructive events. The pathophysiology behind this finding is unknown but may result from unfavorable changes in the upper anatomy due to higher energy delivered to the hypoglossal nerve and surrounding structures. This high energy may also lead to genioglossus muscle fatigue. This finding highlights the importance of hypoglossal nerve stimulation titration with a dedicated sleep study.
Citation: Chang M, Moore V, Eng K, Ryden A, Zeidler M. Hypoglossal nerve stimulation over-titration. . 2024;20(6):999-1001.