Sensitivity and specificity of handheld one lead ECG detecting atrial fibrillation in an outpatient clinic setting.

J Electrocardiol

Karolinska Institutet, Dept of Clinical Sciences, Danderyd University Hospital, Stockholm, Sweden; Dept of Cardiology, Danderyd University Hospital, Stockholm, Sweden.

Published: March 2024

AI Article Synopsis

Article Abstract

Aims: To validate the sensitivity and specificity of the Zenicor One handheld ECG device for detection of atrial fibrillation in an outpatient clinical setting.

Methods And Results: Patients attending outpatient clinics at Danderyd Hospital (n = 220) were examined with one lead handheld ECG immediately after standard care 12‑lead ECG recording. Twelve recordings were excluded (atrial flutter or pacing) or missing. The recordings were dichotomously categorized as "atrial fibrillation" or "not atrial fibrillation" by two senior cardiologists. In cases of diverging interpretations, a third senior cardiologist had the deciding vote. Sensitivity and specificity in diagnosing atrial fibrillation was calculated with 12‑lead ECG as gold standard. Sensitivity and specificity for diagnosis of atrial fibrillation with one lead handheld ECG and 12‑lead ECG as gold standard was 98% and 99% respectively.

Conclusion: In a health-care outpatient setting, Zenicor One handheld ECG had high sensitivity and specificity for detection of atrial fibrillation when compared with 12‑lead ECG.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jelectrocard.2024.02.001DOI Listing

Publication Analysis

Top Keywords

sensitivity specificity
20
atrial fibrillation
20
handheld ecg
16
12‑lead ecg
16
ecg
9
fibrillation outpatient
8
zenicor handheld
8
detection atrial
8
lead handheld
8
ecg gold
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!