Purpose: To assess the accuracy of reused iCare probes after disinfection with 70% isopropyl alcohol and ethylene oxide gas compared to new iCare probes and Goldmann applanation tonometry (GAT).

Design: Prospective comparative analysis.

Participants: A total of 118 eyes from 59 patients recruited from the Aravind Eye Hospital glaucoma clinic in Tirupati, South India.

Methods: Intraocular pressure (IOP) was measured on each eye using a new iCare tonometer probe, an iCare probe previously used and disinfected 1 time prior (once used probe) and 5 times prior (multiply used probe), as well as with GAT. Probes were disinfected after each use with 70% isopropyl alcohol swabs and ethylene oxide sterilization.

Main Outcome Measures: Agreement demonstrated with intraclass correlation coefficients (ICCs), mean difference in IOP values with limits of agreement, and Bland-Altman plots among IOP measurement approaches.

Results: Compared to new iCare probes, both once used probes (ICC = 0.989, 95% confidence interval [CI] 0.985-0.993) and multiply used probes (ICC = 0.989, 95% CI 0.984-0.992) showed excellent agreement, and the mean difference in IOP was minimal for both once used probes (0.70 mmHg, 95% CI 0.29-1.11) and multiply used probes (0.75 mmHg, 95% CI 0.66-0.82) compared to new probes. Bland-Altman plots demonstrated minimal differences between new and reused probes across the spectrum of IOP. When comparing multiply used probes to once used probes, there was a high level of agreement (0.993) (95% CI 0.990-0.995) and negligible mean IOP difference 0.04 mmHg (95% CI 0.32-0.40). Additionally, ICC values for new probes (0.966, 95% CI 0.951-0.976), once used probes (0.958, 95% CI 0.940-0.971), and multiply used probes (0.957, 95% CI 0.938-0.970) compared to GAT were similar and all showed excellent agreement. Both new and reused iCare probes underestimated IOP by 2 to 3 mmHg compared to GAT.

Conclusions: In this prospective comparative analysis, we found that reusing iCare probes up to 5 times does not compromise the accuracy of IOP measurements when disinfected with 70% isopropyl alcohol swabs and ethylene oxide. Reusing iCare probes has the potential to transform care by reducing cost, decreasing environmental waste, and allowing for glaucoma screening camps and increased glaucoma monitoring in low resource settings leading to earlier identification and treatment of glaucoma.

Financial Disclosure(s): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ogla.2024.02.005DOI Listing

Publication Analysis

Top Keywords

icare probes
28
probes
19
multiply probes
16
reused icare
12
70% isopropyl
12
isopropyl alcohol
12
ethylene oxide
12
mmhg 95%
12
icare
9
95%
9

Similar Publications

Background: Bacterial contamination of medical equipment is a significant risk in both human and veterinary medicine, leading to potential cross-contamination between clinicians, technicians, patients, and healthcare professionals. Pathogens can persist on surfaces, resulting in healthcare-associated infections. Tonometry used to measure intraocular pressure for diagnosing conditions like glaucoma and uveitis, is crucial in both human and veterinary care.

View Article and Find Full Text PDF

Purpose: To assess the accuracy of reused iCare probes after disinfection with 70% isopropyl alcohol and ethylene oxide gas compared to new iCare probes and Goldmann applanation tonometry (GAT).

Design: Prospective comparative analysis.

Participants: A total of 118 eyes from 59 patients recruited from the Aravind Eye Hospital glaucoma clinic in Tirupati, South India.

View Article and Find Full Text PDF

Background: Intraoperative measurement of intraocular pressure (IOP) opens up possibilities to investigate the influence of IOP on the stability of the eyeball during the performance of surgical maneuvers. Few techniques have been described, none of which are commercially available. This study investigated the use of sterilized probes for rebound tonometry, which could be used intraoperatively.

View Article and Find Full Text PDF

Oxygen saturation in retinal vessels and their correlation with endothelial microparticles in diabetes mellitus and/or cardiovascular disease.

Microvasc Res

July 2022

Centre for Vision across the Life Span, School of Applied Sciences, University of Huddersfield, Queensgate, HD1 3DH Huddersfield UK; University of Birmingham Institute for Cardiovascular Sciences, City Hospital, Birmingham B18 7QH, UK.

Purpose: Retinal oxygen supply is a critical requirement in ocular function, and when inadequate can lead to retinopathy. Endothelial dysfunction is a leading pathophysiology in diabetes and cardiovascular disease and may be assessed by endothelial microparticles (EMPs). We hypothesised links between retinal vessel oxygenation and EMPs, expecting these indices to be more adverse in those with both DM and CVD.

View Article and Find Full Text PDF

Significance: Intraocular pressure (IOP) profiling is an important component of the glaucoma examination. Two techniques for profiling are the water drinking test (WDT) and iCare HOME phasing, but the correlations between techniques and their ease of deployment have not been studied. These questions are important in determining suitability for clinical deployment.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!