Background And Purpose: In-stent stenosis is commonly observed after stent implantation. There is no consensus on the contributing factors for in-stent stenosis, especially for aneurysms located at or beyond the circle of Willis in the anterior circulation. This study aimed to investigate the morbidity and determinants of in-stent stenosis in distal anterior circulation aneurysms following the implantation of Pipeline Embolization Devices.

Materials And Methods: Patients who underwent Pipeline Embolization Device treatment at our center between January 1, 2018, and June 15, 2023, were enrolled. Distal anterior circulation aneurysms were defined as those occurring at or beyond the circle of Willis, including anterior communicating artery aneurysms, anterior cerebral artery aneurysms, and MCA aneurysms. Baseline information, aneurysm characteristics, and follow-up data of patients were analyzed. Patients were divided into 2 groups: the in-stent stenosis group (patients with a loss of >25% of the lumen diameter of the parent artery) and the non-in-stent stenosis group. Binary logistic regression and restricted cubic spline curves were used to explore risk factors.

Results: We included 85 cases of 1213 patients treated with flow-diverter devices at our hospital. During an average follow-up period of 9.07 months, the complete occlusion rate was 77.64%. The overall incidence of in-stent stenosis was 36.47% (31/85), of which moderate stenosis accounted for 9.41% (8/85), and severe stenosis, 5.88% (5/85) (triglyceride-glucose index ≥ 8.95; OR = 6.883, = .006). The difference in diameters between the stent and parent artery of ≥0.09 mm (OR = 6.534, = .015) and 55 years of age or older (OR = 3.507, = .036) were risk factors for in-stent stenosis. The restricted cubic spline curves indicated that the risk of in-stent stenosis increased as the difference in diameter between stent and parent artery and the triglyceride-glucose index increased.

Conclusions: Compared with the on-label use of Pipeline Embolization Devices, the rate of in-stent stenosis did not obviously increase when treating distal anterior circulation aneurysms with these devices. The incidence of in-stent stenosis was 36.47% when defined as a lumen diameter loss of >25%, and 15.2% when defined as a lumen diameter loss of >50%. Stent-size selection and biochemical indicators can potentially impact the incidence of in-stent stenosis.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11288563PMC
http://dx.doi.org/10.3174/ajnr.A8144DOI Listing

Publication Analysis

Top Keywords

in-stent stenosis
44
anterior circulation
20
pipeline embolization
16
stenosis
14
circle willis
12
distal anterior
12
circulation aneurysms
12
lumen diameter
12
parent artery
12
incidence in-stent
12

Similar Publications

Objective: To explore the prevalence and risk factors of carotid artery (CA) stenosis among subclavian steal syndrome (SSS) patients and to record their prognoses.

Methods: This observational study was retrospective. From January 2015 to October 2022, 169 patients were diagnosed with SSS.

View Article and Find Full Text PDF

Objective: it was to evaluate the efficacy and safety of rapamycin-eluting stents at different doses in the treatment of coronary artery narrowing in miniature pigs.

Methods: a total of 20 miniature pigs were randomly assigned into four groups: S1 group (low-dose rapamycin-coated stent, 55 µg/mm), S2 group (medium-dose rapamycin-coated stent, 120 µg/mm), S3 group (high-dose rapamycin-coated stent, 415 µg/mm), and D0 group (bare metal stent). The stent size was 3.

View Article and Find Full Text PDF

Evaluation of in-stent restenosis after carotid artery stenting with superb microvascular imaging: initial findings.

Acta Radiol

January 2025

Department of Radiology, Interventional Radiology, Sağlık Bilimleri Üniversitesi Ankara Dışkapı Yıldırım Beyazıt Eğitim ve Araştırma Hastanesi, Ankara, Türkiye.

Background: Carotid artery stenting (CAS) is an interventional management in preventing ischemic stroke caused by carotid artery stenosis. After the treatment with CAS, in-stent restenosis caused by neointimal hyperplasia may develop.

Purpose: This study aims to obtain a better determination of neointimal hyperplasia using superb microvascular imaging (SMI), which provides a high-quality visualization of the endoluminal lesions, and to compare these results with B-mode and Doppler ultrasound (US).

View Article and Find Full Text PDF

Introduction: In the past decade, flow diverters (FDs) have increasingly been used to treat cerebral aneurysms with unfavorable morphology in which other endovascular techniques fall short of being as effective. In-stent stenosis (ISS) is one of the most puzzling and frequent risks of flow diversion therapy observed on follow-ups. This complication, although mostly placid in its clinical course, can have dire consequences if patients become symptomatic.

View Article and Find Full Text PDF

Introduction: Coarctation of the aorta is a congenital narrowing of the thoracic aorta associated with hypertension and significant pressure gradients across the coarctation site. Coarctoplasty by percutaneous approach is the preferred method of treatment. However, complications like stent dislodgement may result and must be immediately managed to prevent adverse outcomes.

View Article and Find Full Text PDF