Background: To assess the risk of unsuccessful conversion of ventricular fibrillation during defibrillation testing (DFT) with the subcutaneous implantable cardioverter-defibrillator (S-ICD), the PRAETORIAN score has been proposed.

Objective: The purpose of this study was to validate the PRAETORIAN score in a large S-ICD collective.

Methods: A retrospective single-center analysis of S-ICD patients receiving intraoperative DFT was performed. DFT was performed using a stepwise protocol with 65-J standard polarity, change of polarity, increase to 80 J, and repositioning if necessary. If all DFTs failed, we switched to a transvenous ICD.

Results: Overall, 398 patients were analyzed (268 male [67.3%]; mean age 42.4 ± 15.9 years; mean body mass index [BMI] 25.9 ± 4.8 kg/m). Successful DFT with the first ICD shock was observed in 264 patients (66.3%). One hundred fourteen patients were defibrillated with the second (n = 104) or third (n = 10) DFT after changing shock polarity and/or shock energy. Overall, 20 patients needed at least 3 DFT (ie, 80 J and/or re-positioning). The majority (n = 88 [65.7%]) of DFT failures occurred before 2015 with the first-generation S-ICD. PRAETORIAN score was an independent predictor of DFT failure (odds ratio [OR] 1.007; 95% confidence interval [CI] 1.003-1.011 P ≤.001), while whereas BMI alone was not (P = .31). Presence of hypertrophic cardiomyopathy (HCM) (OR 2.6; 95% CI 1.3-4.4; P = .004) was predictive for at least 1 unsuccessful DFT in our multivariate regression analysis.

Conclusion: PRAETORIAN score proved to be a useful and valid predictive tool for successful DFT, whereas BMI only had a limited role. Patients with HCM were at increased risk for DFT failure or needed higher DFT energy.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2024.02.032DOI Listing

Publication Analysis

Top Keywords

praetorian score
20
dft
12
score large
8
subcutaneous implantable
8
implantable cardioverter-defibrillator
8
s-icd praetorian
8
dft performed
8
successful dft
8
dft failure
8
patients
6

Similar Publications

Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative to traditional ICDs. The PRAETORIAN score, based on chest radiographs, has been validated to predict the probability of successful S-ICD defibrillation testing by assessing factors like fat thickness between the coil and sternum and generator placement.

Objective: This study evaluated the correlation between the PRAETORIAN score and clinical characteristics, as well as implantation variables.

View Article and Find Full Text PDF

Background: Noninferiority of omitting intraoperative defibrillation threshold (DFT) testing has been documented for transvenous implantable cardioverter defibrillators (ICD) whereas data for the subcutaneous-ICD (S-ICD) regarding the need for DFT testing, especially during S-ICD generator replacement, is not available.

Methods: A total of 112 consecutive patients who underwent S-ICD generator replacement and routine testing were included in this retrospective single-center study and analyzed regarding the outcome of intraoperative DFT.

Results: The majority of patients (87.

View Article and Find Full Text PDF

Quality of Life in Subcutaneous or Transvenous Implantable Cardioverter-Defibrillator Patients: A Secondary Analysis of the PRAETORIAN Trial.

Circ Cardiovasc Qual Outcomes

November 2024

Department of Cardiology, Amsterdam Cardiovascular Sciences Heart Failure and Arrhythmias, Amsterdam University Medical Centers Location University of Amsterdam, Heart Center, the Netherlands (R.E.K., J.A.d.V., L.V.A.B., T.F.B., S.P., A.-F.B.E.Q., L.S., W.v.d.S., A.d.W., J.R.d.G., K.M.K., J.G.P.T., A.A.M.W., L.R.A.O.N.).

Article Synopsis
  • The S-ICD was created to eliminate lead-related issues found in the TV-ICD, as it is an external device that sits under the skin rather than using leads inside the body.
  • This analysis comes from the PRAETORIAN trial, where patients were randomly assigned to either S-ICD or TV-ICD and assessed for quality of life through various questionnaires at different stages.
  • Results showed no significant differences in physical and mental well-being between the groups, but patients who experienced a shock recently reported lower social functioning and emotional health compared to those who did not.
View Article and Find Full Text PDF

Effect of posture on position of subcutaneous implantable cardioverter-defibrillator.

Heart Rhythm

August 2024

Department of Cardiology, Heart Center, Amsterdam Cardiovascular Sciences Heart Failure & Arrhythmias, Amsterdam UMC, Amsterdam, The Netherlands.

Background: The PRAETORIAN score was developed as an alternative for defibrillation testing after subcutaneous implantable cardioverter-defibrillator implantation to assess 3 aspects of implant position on a bidirectional chest radiograph. The score is validated on a standard standing chest radiograph with arms elevated in the lateral view.

Objective: We aimed to evaluate the effect of different anatomic positions on the PRAETORIAN score.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!