Validity and reliability of the Oura Ring Generation 3 (Gen3) with Oura sleep staging algorithm 2.0 (OSSA 2.0) when compared to multi-night ambulatory polysomnography: A validation study of 96 participants and 421,045 epochs.

Sleep Med

Precision Health, Department of Bioengineering, Graduate School of Engineering, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656, Japan; Department of Clinical Sciences, Lund University, Skåne University Hospital, Malmö, Sweden; Department of Diabetes and Metabolic Diseases, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-0033, Japan.

Published: March 2024

Purpose: To evaluate the validity and the reliability of the Oura Ring Generation 3 (Gen3) with Oura Sleep Staging Algorithm 2.0 (OSSA 2.0) through multi-night polysomnography (PSG).

Participants And Methods: Participants were 96 generally healthy Japanese men and women aged between 20 and 70 years contributing with 421,045 30-s epochs. Sleep scoring was performed according to American Academy of Sleep Medicine criteria. Each participant could contribute with a maximum of three polysomnography (PSG) nights. Within-participant means were created for each sleep measure and paired t-tests were used to compare equivalent measures obtained from the PSG and Oura Rings (non-dominant and dominant hand). Agreement between sleep measures were assessed using Bland-Altman plots. Interrater reliability for epoch accuracy was determined by prevalence-adjusted and bias-adjusted kappa (PABAK).

Results: The Oura Ring did not significantly differ from PSG for the measures time in bed, total sleep time, sleep onset latency, sleep period time, wake after sleep onset, time spent in light sleep, and time spent in deep sleep. Oura Rings worn on the non-dominant- and dominant-hand underestimated sleep efficiency by 1.1 %-1.5 % and time spent in REM sleep by 4.1-5.6 min. The Oura Ring had a sensitivity of 94.4 %-94.5 %, specificity of 73.0 %-74.6 %, a predictive value for sleep of 95.9 %-96.1 %, a predictive value for wake of 66.6 %-67.0 %, and accuracy of 91.7 %-91.8 %. PABAK was 0.83-0.84 and reliability was 94.8 %. Sleep staging accuracy ranged between 75.5 % (light sleep) and 90.6 % (REM sleep).

Conclusions: The Oura Ring Gen3 with OSSA 2.0 shows good agreement with PSG for global sleep measures and time spent in light and deep sleep.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.sleep.2024.01.020DOI Listing

Publication Analysis

Top Keywords

oura ring
20
sleep
19
time spent
16
sleep staging
12
oura
9
validity reliability
8
reliability oura
8
ring generation
8
generation gen3
8
gen3 oura
8

Similar Publications

Background And Aims: Wearable devices capture physiological signals non-invasively and passively. Many of these parameters have been linked to inflammatory bowel disease (IBD) activity. We evaluated the associative ability of several physiological metrics with IBD flares and how they change before the development of flare.

View Article and Find Full Text PDF

Deriving Accurate Nocturnal Heart Rate, rMSSD and Frequency HRV from the Oura Ring.

Sensors (Basel)

November 2024

Centre for Sleep and Cognition, Yong Loo Lin School of Medicine, National University of Singapore (NUS), Singapore 117549, Singapore.

Cardiovascular diseases are a major cause of mortality worldwide. Long-term monitoring of nighttime heart rate (HR) and heart rate variability (HRV) may be useful in identifying latent cardiovascular risk. The Oura Ring has shown excellent correlation only with ECG-derived HR, but not HRV.

View Article and Find Full Text PDF

This dataset was collected from university students before, during, and after the COVID-19 lockdown in Southern California. Data collection happened continuously for the average of 7.8 months (=3.

View Article and Find Full Text PDF

Sleep tracking by consumers is becoming increasingly prevalent; yet, few studies have evaluated the accuracy of such devices. We sought to evaluate the accuracy of three devices (Oura Ring Gen3, Fitbit Sense 2, and Apple Watch Series 8) compared to the gold standard sleep assessment (polysomnography (PSG)). Thirty-five participants (aged 20-50 years) without a sleep disorder were enrolled in a single-night inpatient study, during which they wore the Oura Ring, Fitbit, and Apple Watch, and were monitored with PSG.

View Article and Find Full Text PDF

During the COVID-19 pandemic, caregiving responsibilities may have been associated with increased substance use. To characterize substance use to cope with stress and willingness to seek help among (i) parents, (ii) unpaid caregivers of adults, and (iii) parent-caregivers. Data were analyzed for 10,444 non-probabilistic internet-based survey respondents of the COVID-19 Outbreak Public Evaluation (COPE) initiative (5227 females, 5217 males).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!