Background: The effects of awake prone positioning (APP) on respiratory mechanics in patients with COVID-19 are not well characterized. The aim of this study was to investigate changes of global and regional lung volumes during APP compared with the supine position using electrical lung impedance tomography (EIT) in patients with hypoxemic respiratory failure due to COVID-19.

Materials And Methods: This exploratory non-randomized cross-over study was conducted at two university hospitals in Sweden between January and May 2021. Patients admitted to the intensive care unit with confirmed COVID-19, an arterial cannula in place, a PaO2/FiO2 ratio <26.6 kPa (<200 mmHg) and high-flow nasal oxygen or non-invasive ventilation were eligible for inclusion. EIT-data were recorded at supine baseline, at 30 and 60 minutes after APP-initiation, and 30 minutes after supine repositioning. The primary outcomes were changes in global and regional tidal impedance variation (TIV), center of ventilation (CoV), global and regional delta end-expiratory lung-impedance (dEELI) and global inhomogeneity (GI) index at the end of APP compared with supine baseline. Data were reported as median (IQR).

Results: All patients (n = 10) were male and age was 64 (47-73) years. There were no changes in global or regional TIV, CoV or GI-index during the intervention. dEELI increased from supine reference value 0 to 1.51 (0.32-3.62) 60 minutes after APP (median difference 1.51 (95% CI 0.19-5.16), p = 0.04) and returned to near baseline values after supine repositioning. Seven patients (70%) showed an increase >0.20 in dEELI during APP. The other EIT-variables did not change during APP compared with baseline.

Conclusion: Awake prone positioning was associated with a transient lung recruiting effect without changes in ventilation distribution measured with EIT in patients with hypoxemic respiratory failure due to COVID-19.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10880988PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0299199PLOS

Publication Analysis

Top Keywords

lung impedance
8
awake prone
8
prone positioning
8
non-randomized cross-over
8
cross-over study
8
impedance changes
4
changes awake
4
positioning covid-19
4
covid-19 non-randomized
4
study background
4

Similar Publications

Aerodynamic and Acoustic Power in Infant Cry.

J Voice

January 2025

Utah Center for Vocology, University of Utah, Salt Lake City, UT; National Center for Voice and Speech, Salt Lake City, UT. Electronic address:

Objectives: Acoustic and aerodynamic powers in infant cry are not scaled downward with body size or vocal tract size. The objective here was to show that high lung pressures and impedance matching are used to produce power levels comparable to those in adults.

Study Design And Methodology: A computational model was used to obtain power distributions along the infant airway.

View Article and Find Full Text PDF

[Effect of extra corporeal reducing pre-load on pulmonary mechanical power in patients with acute respiratory distress syndrome].

Zhonghua Wei Zhong Bing Ji Jiu Yi Xue

December 2024

Department of Public Utilities Development, Henan Provincial People's Hospital, People's Hospital of Zhengzhou University, Zhengzhou 450003, Henan, China.

Objective: To explore the effects of veno-venous extra corporeal carbon dioxide removal (V-V ECCOR) on local mechanical power and gas distribution in the lungs of patients with mild to moderate acute respiratory distress syndrome (ARDS) receiving non-invasive ventilation.

Methods: Retrospective research methods were conducted. Sixty patients with mild to moderate ARDS complicated with renal insufficiency who were transferred to the respiratory intensive care unit (RICU) through the 96195 platform critical care transport green channel from January 2018 to January 2020 at the collaborative hospitals of Henan Provincial People's Hospital were enrolled.

View Article and Find Full Text PDF

Dynamic EIT technology for real-time non-invasive monitoring of acute pulmonary embolism: a porcine model experiment.

Respir Res

January 2025

Shaanxi Provincial Key Laboratory of Bioelectromagnetic Detection and Intelligent Perception, Department of Biomedical Engineering, Air Force Medical University, Xi'an, 710032, China.

Background: Acute pulmonary embolism represents the third most prevalent cardiovascular pathology, following coronary heart disease and hypertension. Its untreated mortality rate is as high as 20-30%, which represents a significant threat to patient survival. In view of the current lack of real-time monitoring techniques for acute pulmonary embolism, this study primarily investigates the potential of the pulsatility electrical impedance tomography (EIT) technique for the detection and real-time monitoring of acute pulmonary embolism through the collection and imaging of the pulsatile signal of pulmonary blood flow.

View Article and Find Full Text PDF

Background/aim: Chronic obstructive pulmonary disease (COPD) is often complicated by sarcopenia, a condition of reduced muscle mass and function that adversely affects quality of life, lung function, and exacerbation rates. Ultrasonography could be an effective tool for detecting sarcopenia, notably by assessing diaphragmatic function, which may indicate muscle health in COPD patients. This study aims to evaluate the effectiveness of diaphragmatic ultrasound in detecting sarcopenia among COPD patients.

View Article and Find Full Text PDF

Positive end-expiratory pressure (PEEP) titration is crucial for improving oxygenation and preventing ventilator-induced lung injury in acute hypoxemic respiratory failure. Electrical impedance tomography (EIT) offers real-time, bedside monitoring of lung ventilation distribution, potentially guiding individualized PEEP settings.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!