Two different autorefractors for vision screening in children and adolescents.

Int J Ophthalmol

National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou 325027, Zhejiang Province, China.

Published: February 2024

Aim: To compare the consistency of two autorefractors (Tianle RM-9000 and Topcon KR-800) for school-age myopia children, and to provide a basis for largescale data analysis and comparison.

Methods: The refractive error in 909 subjects (age 4-18y) were measured using both autorefractors without cycloplegia. The data were analyzed using Fourier decomposition and the correlation coefficients, intraclass correlation coefficients (ICC), and Bland-Altman limits of agreement (LoA) for each parameter were calculated.

Results: There was a strong correlation between the spherical equivalent (SE), sphere diopter (DS), and cylinder diopter (DC) readings of the Tianle RM-9000 and those of the Topcon KR-800, with correlation coefficient values of 0.98, 0.98 and 0.83 and ICC values of 0.99, 0.99 and 0.93, respectively. However, the correlation coefficients and ICC values of J and J were unreliable (=-0.004, -0.034; both ICC<0.10). Bland-Altman analysis revealed that SE, DS, and DC measured by the Tianle RM-9000 were significantly biased toward myopia compared with the Topcon KR-800, and the mean differences were -0.072, -0.026, -0.091 D, respectively (all <0.01). The minimum absolute value of the difference within the 95% LoA for SE, DS, and DC was 0.63 D, 0.50 D, 0.62 D, respectively; all these values were in the clinically acceptable range. For J and J, the mean differences were close to zero (=0.43, 0.84); however, the 95% LoA were relatively wide (J SD: 0.53; 95%CI: -1.00, 1.10; J SD: 0.52; 95%CI: -1.00, 1.00).

Conclusion: The two autorefractors are consistent with each other, as the differences in SE, DS, and DC were within the clinically acceptable range. Readers can compare the data measured by either device in different studies and use the two devices in the same study to generate a dataset that can be analyzed together. However, the J and J vectors are unreliable and should not be used to assess astigmatism.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10827628PMC
http://dx.doi.org/10.18240/ijo.2024.02.16DOI Listing

Publication Analysis

Top Keywords

correlation coefficients
12
tianle rm-9000
8
rm-9000 topcon
8
topcon kr-800
8
coefficients icc
8
icc values
8
correlation
5
autorefractors vision
4
vision screening
4
screening children
4

Similar Publications

Validation of lactate and glucose in cerebrospinal fluid (CSF) on a Radiometer blood gas analyzer ABL90 Flex plus.

Clin Biochem

January 2025

Pathology and Laboratory Medicine Program, Health Sciences Centre, St. John's, Newfoundland and Labrador, Canada; Memorial University of Newfoundland, Health Sciences Centre, St. John's, Newfoundland and Labrador, Canada. Electronic address:

Purpose: Rapid determination of cerebrospinal fluid. (CSF) glucose and lactate is required by emergency rooms and intensive care units. Long turnaround time (TAT) on test results negatively impacts timely diagnosis and treatment of neurological infections like meningitis.

View Article and Find Full Text PDF

Metoprolol vs diltiazem for atrial fibrillation with rapid ventricular rate: Systematic review and meta-analysis of adverse events.

Am J Emerg Med

December 2024

Department of Emergency Medicine, University of Maryland School of Medicine, Baltimore, MD, USA; Departments of Pharmacy and Emergency Medicine, University of Rochester Medical Center, Rochester, New York, USA. Electronic address:

Background: Intravenous (IV) diltiazem and metoprolol are commonly used to achieve rate control for atrial fibrillation with RVR (Afib with RVR), and are both recommended as first-line by current guidelines. While prior studies investigated the efficacy of these medications, there is little evidence available regarding the risk of adverse events (AEs) with their use.

Methods: We identified randomized controlled trials (RCT) and observational studies reporting rates of AEs following administration of IV diltiazem and metoprolol for Afib with RVR by searching PubMed, SCOPUS, EMBASE, and Cochrane Library.

View Article and Find Full Text PDF

Rationale And Objectives: Accurate assessment of hip morphology is crucial for the diagnosis and management of hip pathologies. Traditional manual measurements are prone to mistakes and inter- and intra-reader variability. Artificial intelligence (AI) could mitigate such issues by providing accurate and reproducible measurements.

View Article and Find Full Text PDF

How does globally accumulated tropical cyclone energy vary in response to a changing climate?

Sci Bull (Beijing)

December 2024

Department of Ocean Science and Engineering, Southern University of Science and Technology, Shenzhen 518055, China. Electronic address:

How tropical cyclone (TC) activity varies in response to a changing climate is widely debated. The accumulated cyclone energy (ACE) is one of the indicators of TC activity and has attracted considerable attention because of its close relationship with the damages caused by TCs. Previous studies have focused on detecting long-term trends in global ACE; however, the results are inconclusive.

View Article and Find Full Text PDF

Strength and endurance of the lumbar extensor muscles and their predictors: A cross-sectional study in healthy subjects.

J Electromyogr Kinesiol

December 2024

Department of Neurology, Center for Neuromuscular Diseases (Associated National Center in the European Reference Network ERN EURO-NMD), University Hospital Brno, Brno, Czechia; Faculty of Medicine, Masaryk University, Brno, Czechia; Department of Rehabilitation, University Hospital Brno, Brno, Czechia; Department of Public Health, Faculty of Medicine, Masaryk University, Brno, Czechia.

The purpose of this cross-sectional study was to present the values of maximal isometric lumbar extensor muscle strength (MILEMS) and lumbar extensor muscle endurance (LEME) in healthy subjects and to study the influence of physiological factors on these parameters. MILEMS using a hand-held dynamometer in three positions (prone, sitting, standing) and LEME using the Biering-Sørensen test were assessed in 115 healthy volunteers. The MILEMS measurements for the specific positions were correlated with each other (Pearson correlation coefficients 0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!