Purpose: To elucidate the dosimetric errors caused by a model-based algorithm in lung stereotactic body radiation therapy (SBRT) with Helical TomoTherapy (HT) using Monte Carlo (MC)-based dose verification software.

Methods: For 38 plans of lung SBRT, the dose calculation accuracy of a treatment planning system (TPS) of HT was compared with the results of DoseCHECK, the commercial MC-based independent verification software. The following indices were extracted to evaluate the correlation of dosimetric errors: (1) target volume, (2) average computed tomography (CT) value of the planning target volume (PTV) margin, and (3) average CT value of surrounding 2-mm area of the PTV (PTV ring). Receiver operating characteristic (ROC) analyses determined the threshold for 5% of differences in PTV D. Then, the 38 plans were classified into two groups using the cutoff values of ROC analysis for these three indices. Dosimetric differences between groups were statistically compared using the Mann-Whitney U test.

Results: TPS of HT overestimated by more than 5% in the PTV D in 16 of 38 plans. The PTV ring showed the strongest correlation with dosimetric differences. The cutoff value for the target volume, the PTV margin, and the PTV ring was 14.7 cc, -754 HU, and -708 HU, respectively. The area under the curve (AUC) for the target volume, the PTV margin, and the PTV ring were 0.835, 0.878, and 0.932, respectively. Dosimetric errors more than 5% were observed when the PTV volume was less than 15 cc or when the CT value around the target was less than -700 HU.

Conclusion: The TPS of HT might overestimate the PTV dose by more than 5% if any the three indices in this study were below threshold. Therefore, independent verification with an MC-based algorithm should be strongly recommended for lung SBRT in HT.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11087163PMC
http://dx.doi.org/10.1002/acm2.14305DOI Listing

Publication Analysis

Top Keywords

target volume
16
ptv ring
16
lung sbrt
12
dosimetric errors
12
ptv
12
volume ptv
12
ptv margin
12
helical tomotherapy
8
dose verification
8
verification software
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!