Introduction: Reverse osmosis (RO) membrane, key component of water-purifying equipment, is often stored in protection fluid containing substances such as glycerol, which may contaminate the water at replacement. This study aims to explore the effects of RO membrane replacement on clinical chemistry and immunoassay, particularly triglyceride (TG), providing reference for managing test interference caused by RO membrane replacement.
Materials And Methods: The RO membrane of water-purifying equipment A, which provided water to C16000 biochemistry analyzer (Abbott Laboratories, Abbott Park, USA) and E801 electrochemiluminescence analyzer (Roche, Basel, Switzerland), was replaced. Water resistivity was recorded, and quality control (QC) tests were performed on C16000 and E801. Moreover, TG was measured in 29 of selected serum samples on C16000 at 0.5h and 10.5h after RO membrane replacement and on reference biochemistry analyzer BS2000M (Mindray Biomedical Electronics Co., Shenzhen, China), which was connected to water-purifying equipment B without RO membrane replacement. Finally, blank, calibrator 1 and calibrator 2 of TG reagent were measured on C16000 before and at 0.5h, 2.5h and 10.5h after RO membrane replacement. All statistical analyses of data were done using GraphPad Prism (GraphPad Software Inc., San Diego, USA), and a value of P < 0.05 was considered statistically significant.
Results: After RO membrane replacement, all QC results of clinical chemistry and immune tests passed except TG that showed positive bias of 536% and 371% at two levels, respectively. Moreover, TG results of the same serum samples were significantly higher at 0.5h than 10.5h after RO membrane replacement. Meanwhile, there was worse agreement and correlation of TG results between C16000 and BS2000M at 0.5h than 10.5h after replacement. Furthermore, the absorbance of TG blank, calibrator 1 and calibrator 2 was significantly higher at 0.5h and 2.5h after replacement than before replacement, and the absorbance gradually returned to normal value at 10.5h after replacement.
Conclusions: Replacement of RO membrane could cause significant interference to TG test while have no effects on other laboratory tests performed in the study, which may be due to glycerol contamination. Our data provides important reference for management of test interference caused by RO membrane replacement. Clinical laboratory should observe the effects of RO membrane replacement on laboratory tests through both water quality monitoring and QC detection.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10864026 | PMC |
http://dx.doi.org/10.11613/BM.2024.010705 | DOI Listing |
Multimed Man Cardiothorac Surg
January 2025
Department of Cardiothoracic Surgery, Faculty of Medicine, Mansoura University Mansoura, Egypt.
The Ross procedure continues to be the best procedure to address unrepairable aortic valve pathology, especially in young adults. The Achilles heel of this procedure has been aortic root dilation and the potential need for a reoperation that may be associated with slightly increased risks in addition to the need for intervention on the pulmonary outflow tract. Modifying the Ross procedure by autograft inclusion inside a Dacron graft seems to have the potential advantage of stabilizing the autograft diameter, which may be associated with improved durability and decrease the need for future intervention.
View Article and Find Full Text PDFJ Cardiovasc Dev Dis
December 2024
Department of Cardiothoracic and Vascular Surgery, Montefiore Medical Center, Bronx, NY 10467, USA.
The use of extracorporeal membrane oxygenation (ECMO) has emerged as a rescue intervention for hemodynamically unstable patients and prophylactic intraprocedural hemodynamic support in the cardiac catheterization laboratory. The prompt initiation of ECMO provides immediate hemodynamic support and allows for the completion of bridging and/or life-saving interventions. However, there are no clinical practice guidelines for the use of extracorporeal support in this area.
View Article and Find Full Text PDFDent J (Basel)
January 2025
Department of Conservative Dentistry with Endodontics, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, 41-902 Bytom, Poland.
Traditional root canal therapy (RCT) effectively removes diseased or necrotic pulp tissue and replaces it with inorganic materials. Regenerative endodontics is an alternative to conventional RCT by using biologically based approaches to restore the pulp-dentin complex. This review explores emerging techniques, including autogenic and allogenic pulp transplantation, platelet-rich fibrin, human amniotic membrane scaffolds, specialized pro-resolving mediators, nanofibrous and bioceramic scaffolds, injectable hydrogels, dentin matrix proteins, and cell-homing strategies.
View Article and Find Full Text PDFAm J Case Rep
January 2025
Department of Anesthesiology and Intensive Care, Legionowo Hospital, Military Institute of Medicine - National Research Institute, Warsaw, Poland.
BACKGROUND Extracorporeal blood purification strategies were recently developed as adjunctive treatments for sepsis. CytoSorb® is an approved medical device designed to reduce blood levels of inflammatory cytokines. The oXiris high-adsorption membrane filter is used in continuous hemofiltration adsorption.
View Article and Find Full Text PDFFront Neural Circuits
January 2025
Department of Neurobiology, Hokkaido University Graduate School of Medicine, Sapporo, Japan.
A strong repetitive stimulus can occasionally enhance axonal excitability, leading to the generation of afterdischarge. This afterdischarge outlasts the stimulus period and originates either from the physiological spike initiation site, typically the axon initial segment, or from ectopic sites for spike generation. One of the possible mechanisms underlying the stimulus-induced ectopic afterdischarge is the local depolarization due to accumulated potassium ions surrounding the axonal membranes of the distal portion.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!