Objective: To investigate the aplication of low contrast agent concentration and low tube voltage in coronary CTA on patients with high BMI (26kg/m
Methods: 60 patients with high BMI (26kg/m2
Result: There was no significant difference in subjective scores of coronary artery image quality between the two high BMI subgroups (P>0.05).The ED of group A (1.103±0.101 mSv) was significantly lower than that of group a (4.663±0.412 mSv) (P<0.001).There was no significant difference in mean CT value, SNR and CNR between the two subgroups (P>0.05).The image excellent index (FOM) of group A was higher than that of group a (P<0.05).The total iodine content and iodine injection rate in group A were lower than those in group a (P<0.001).The difference of subjective scores of coronary artery image quality between the two ultra-high subgroups was significant. The mean CT value, SNR and CNR of group B were lower than those of group b (P<0.05). The images of 14 patients in group B could not meet the diagnosis demand.
Conclusion: It is feasible to reduce the tube voltage to 70KV in patients with abdominal BMI with high BMI (26Kg/m
Download full-text PDF |