Across five studies, we present the preliminary technical validation of an infant-wearable platform, LittleBeats™, that integrates electrocardiogram (ECG), inertial measurement unit (IMU), and audio sensors. Each sensor modality is validated against data from gold-standard equipment using established algorithms and laboratory tasks. Interbeat interval (IBI) data obtained from the LittleBeats™ ECG sensor indicate acceptable mean absolute percent error rates for both adults (Study 1, = 16) and infants (Study 2, = 5) across low- and high-challenge sessions and expected patterns of change in respiratory sinus arrythmia (RSA). For automated activity recognition (upright vs. walk vs. glide vs. squat) using accelerometer data from the LittleBeats™ IMU (Study 3, = 12 adults), performance was good to excellent, with smartphone (industry standard) data outperforming LittleBeats™ by less than 4 percentage points. Speech emotion recognition (Study 4, = 8 adults) applied to LittleBeats™ versus smartphone audio data indicated a comparable performance, with no significant difference in error rates. On an automatic speech recognition task (Study 5, = 12 adults), the best performing algorithm yielded relatively low word error rates, although LittleBeats™ (4.16%) versus smartphone (2.73%) error rates were somewhat higher. Together, these validation studies indicate that LittleBeats™ sensors yield a data quality that is largely comparable to those obtained from gold-standard devices and established protocols used in prior research.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10857055 | PMC |
http://dx.doi.org/10.3390/s24030901 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!