The Poly acrylic acid/MIL-88(Fe)-NH composite material, carefully prepared, is employed as a sorbent for the stir bar. The best formula of the composite was selected by investigation of two parameters including the cross-linker of PAA and MIL-88(Fe)-NH content. The prepared stir bar was used for extraction of 2-pentanone, 2-heptanone, ethyl propionate, para-xylene, 1,2,4-trimethylbenzene, o-cresol, m-cresol in urine samples as breast cancer biomarkers with gas chromatography-flame ionization detector. The prepared Poly acrylic acid / MIL-88(Fe)-NH as sorbent for the stir bar demonstrate good repeatability of one bar (relative standard deviation (RSD%) < 4.61 %) and satisfactory reproducibility between two bars (RSD% < 6.85 %). The central composite design method was applied for the optimization of extraction parameters. Under the optimum conditions, linear dynamic ranges for compounds were in the acceptable range with correlation coefficients higher than 0.99. Detection limits of them were less than 1.71 µg L.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.chroma.2024.464708 | DOI Listing |