Background: Diagnostic reference levels (DRLs) are an important metric in identifying abnormally high radiation doses in diagnostic examinations. National DRLs for CT colonography do not currently exist in South Africa, but there are efforts to collect data for a national DRL project.

Objectives: This study investigated radiation doses for CT colonography in adult patients at a large tertiary hospital in South Africa with the aim of setting local DRLs.

Method: Patient data from two CT scanners (Philips Ingenuity and Siemens Somatom go.Top) in the period March 2020 - March 2023 were obtained from the hospital's picture archiving and communication system (PACS) ( = 115). Analysis involved determining the median computed tomography dose index-volume (CTDI) and dose-length product (DLP) values. The findings were compared with DRLs established internationally.

Results: Ingenuity median CTDI was 20 mGy and DLP was 2169 mGy*cm; Somatom median CTDI was 6 mGy and DLP was 557 mGy*cm. Ingenuity exceeded the United Kingdom's (UK) recommended DRLs by 82% and 214%, respectively. Somatom median CTDI and DLP were 45% and 19% lower than UK NDRLs.

Conclusion: Somatom's tin filter and other dose reduction features provided significant dose reduction. These data were used to set DRLs for CT colonography at the hospital; CTDI: 6 mGy and DLP: 557 mGy*cm.

Contribution: In addition to informing radiation protection practices at the level of the institution, the established local DRLs contribute towards implementing regional and national DRLs.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10839235PMC
http://dx.doi.org/10.4102/sajr.v28i1.2809DOI Listing

Publication Analysis

Top Keywords

median ctdi
12
ctdi mgy
12
mgy dlp
12
diagnostic reference
8
reference levels
8
tertiary hospital
8
radiation doses
8
national drls
8
drls colonography
8
south africa
8

Similar Publications

Clinical diagnostic reference levels and image quality metrics for CT in oncology patients.

SA J Radiol

October 2024

Department of Clinical Sciences, Faculty of Health and Environmental Sciences, Central University of Technology, Bloemfontein, South Africa.

Background: In CT, the volumetric CT dose index (CTDI), dose-length product (DLP) and patient's size-specific dose estimates (SSDE) are used as diagnostic reference level (DRL) metrics.

Objectives: To develop clinical local DRL values for CT chest-abdomen-pelvis (CAP) examinations using the CTDI, DLP and SSDE, and to determine the image quality achieved.

Method: In total, 201 cancer patients were included in the study.

View Article and Find Full Text PDF

In 2020, recording and management of exposure doses for computed tomography (CT) became mandatory. In terms of dose management, information regarding the imaged body part is particularly important information for proper tabulation. However, the actual body part to be imaged and the body part information obtained from the device may differ.

View Article and Find Full Text PDF

Image quality and radiation doses in abdominal CT: A multicenter study.

Eur J Radiol

September 2024

Department of Life Sciences and Health, Faculty of Health Science, Oslo Metropolitan University Oslo, Norway; Centre for Research and Innovation, Sunnaas Rehabilitation Hospital, Bjornemyr, Norway. Electronic address:

Purpose: To benchmark image quality and corresponding radiation doses for acute abdominal CT examination across different laboratories and CT manufacturers.

Method: An anthropomorphic phantom was scanned once with local abdominal CT protocols at 40 CT scanners, from four vendors, in thirty-three sites. Quantitative image quality was evaluated by CNR and SNR in the liver and kidney parenchyma.

View Article and Find Full Text PDF

Purpose: To describe the first experience of patient dose optimization in developing AQD, SSDE and image quality-related DRLs for common CT examinations in the adult age group using the concept of AQD.

Materials And Methods: The recent published IQSC from 0 to 4 were used by radiologists for the assessment of image quality. The entire data were collected for five types (brain CT, chest CT, chest HRCT, abdomen KUB CT and abdomen + pelvic CT) CT investigations based on anatomic region (head, chest and abdomen + pelvic).

View Article and Find Full Text PDF

Proposing Multiregional Diagnostic Reference Levels for Common CT Angiography Examinations in Saudi Arabia.

Diagnostics (Basel)

July 2024

Department of Radiological Sciences, College of Applied Medical Sciences, King Khalid University, Abha 62521, Saudi Arabia.

Objectives: Diagnostic reference levels (DRLs) are crucial tools for optimizing radiation exposure during different radiological examinations. This study aimed to establish preliminary DRLs for commonly performed computed tomographic angiography (CTA) examinations in Saudi Arabia.

Methods: Data for three types of CTA examinations (cerebral, pulmonary, and lower-extremity) were collected from six medical cities across Saudi Arabia.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!