Objective: To observe the effect of using hydrogen peroxide in periodic disinfection combining with continuous disinfection of dental unit waterlines and to provide references for the selection of waterway disinfection measures.
Methods: A total of 4 dental units in a hospital of stomatology were selected through convenience sampling. The dental unit waterlines received periodic disinfection once every 4 weeks in addition to continuous disinfection (When the dental units were not used for more than 3 days, an additional periodic disinfection would be performed.). Periodic disinfection referred to filling up the waterlines with a disinfectant solution (1.4% hydrogen peroxide) by using the waterline disinfection device that came with the dental unit, immersing for 24 hours, and then emptying out the disinfectant solution. Continuous disinfection referred to using hydrogen peroxide at a concentration of 0.014% as dental treatment water and using it to flush the waterlines for 2 minutes before any dental treatment in the morning and to flush the waterlines for 30 seconds after each dental treatment. The study lasted for 25 weeks, with periodic disinfection being performed for 7 times and continuous disinfection carried out for the rest of the dental treatment time. During the 25 weeks, water samples were collected from air/water syringes and high-speed handpieces. Then, the water samples were incubated and the bacterial concentration and the qualification rates were calculated accordingly. When the bacterial concentration≤100 CFU/ mL, the water samples were considered to be qualified. Waterline tubes of 1 cm were collected before and after the 25 weeks of disinfection with hydrogen peroxide. Biofilms in the waterline tube were observed under scanning electron microscope.
Results: A total of 352 water samples were collected. Eight water samples were collected before disinfection with hydrogen peroxide, with the median of bacterial concentration being 3140 CFU/mL. On the first day of disinfection with hydrogen peroxide, the median bacterial concentration in dental treatment water was 7.5 CFU/mL. There was a significant difference between the bacterial concentration of the water samples before the disinfection and that after the disinfection (=0.012). A total of 344 water samples were collected after the disinfection, with the median bacterial concentrations for air/water syringes and high-speed handpieces being 11 CFU/mL and 11CFU/mL and the qualified rates being 83.7% and 82.0%, respectively. There was no significant difference in bacterial concentration or the qualification rates. During week 1 through week 9 of the disinfection, the qualification rates of the dental treatment water always exceeded 80% in 8 weeks, with week 3 being the exception. In the two four-week disinfection periods of week 14 through week 17 and week 18 through week 21, the qualification rate was maintained at above 80% for only the first two weeks and started to decrease from the third week. Biofilm morphology was observed under scanning electron microscope. Before the disinfection, the biofilm was found to be a dense structure and the mixture of a large number of bacteria. After 25 weeks of the disinfection, the biofilm structure appeared to be loose and did not show consistent characteristics of a large number of bacteria retained.
Conclusion: Periodic disinfection combined with continuous disinfection using hydrogen peroxide can effectively control contamination in dental unit waterlines. But the cycles of periodic disinfection and the concentration of hydrogen peroxide for continuous disinfection should be further discussed according to the actual clinical situation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10839502 | PMC |
http://dx.doi.org/10.12182/20240160210 | DOI Listing |
Mikrochim Acta
January 2025
College of Medical Technology, Chengdu University of Traditional Chinese Medicine, Chengdu, 611137, China.
An innovative colorimetric sensing strategy was developed for the detection of glucose by the integration of glucose aptamer, glucose oxidase (GOx), and horseradish peroxidase (HRP), termed aptamer proximal enzyme cascade reactions (APECR). In the presence of glucose, aptamer binding enables GOx to catalyze glucose oxidation into HO efficiently. Subsequently, the adjacent HRP catalyzes the oxidation of the peroxidase substrate, 2,2'-biazobis(3-ethylbenzothiazoline-6-sulfonic acid) (ABTS), utilizing the generated HO, resulting in a distinct color change.
View Article and Find Full Text PDFClin Cosmet Investig Dent
January 2025
College of Dentistry, King Saud University, Riyadh, Saudi Arabia.
Aim: This study evaluated the color stability and surface roughness of two universal-shade compared to two nanohybrid composites after staining and external bleaching with 40% hydrogen peroxide.
Methods: Two universal shade resin-based composites and two nanohybrid composites were tested. Twenty disc-shaped specimens from each material were fabricated and divided into two subgroups: one group was stained and bleached (staining group) and the other received bleaching treatment only (control group).
Toxicol Res
January 2025
Department of Applied Chemistry, Dong-Eui University, Busan, 614-714 Republic of Korea.
Imatinib mesylate is a targeted anti-cancer drug with skin pigmentation as a side effect. The action mechanism of imatinib mesylate on melanogenesis remains unclear. The purpose of this study was to elucidate the mechanism of imatinib mesylate on melanogenesis associated with the microphthalmia-associated transcription factor (MITF) signaling pathway in murine melanoma cells.
View Article and Find Full Text PDFACS Appl Mater Interfaces
January 2025
Department of Mechanical and Energy Engineering, Southern University of Science and Technology, Shenzhen 518055, China.
The electrocatalytic conversion of oxygen to hydrogen peroxide offers a promising pathway for sustainable energy production. However, the development of catalysts that are highly active, stable, and cost-effective for hydrogen peroxide synthesis remains a significant challenge. In this study, a novel polyacid-based metal-organic coordination compound (Cu-PW) was synthesized using a hydrothermal approach.
View Article and Find Full Text PDFLangmuir
January 2025
State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, 5625 Renmin Street, Changchun, Jilin 130022, P R China.
The detection of hydrogen peroxide (HO) at elevated concentrations while eliminating oxygen interference presents a significant challenge. Nickel-based stainless steel, such as Hastelloy G35, has shown excellent corrosion resistance. However, it has never been used in electroanalysis.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!