Effect of Restoration Design on the Removal Torque Loss of Implant-supported Crowns after Cyclic Loading.

J Contemp Dent Pract

Department of Prosthodontics, Faculty of Dentistry, Chulalongkorn University, Pathumwan, Bangkok, Thailand, Phone: +66945599524, e-mail: Orcid: https://orcid.org/0000-0002-0660-162X.

Published: December 2023

Aim: To compare the removal torque loss (RTL) percentage of screw-retained, cement-retained, and combined screw- and cement-retained implant-supported crowns after cyclic loading and measure the impact of cyclic loading on removal torque.

Materials And Methods: Thirty-two dental implants (4.0 × 10 mm) in resin blocks and abutments were divided into four groups ( = 8) based on restoration design: combined screw- and cement-retained group (SC), two cement-retained groups: cemented with adhesive resin cement (AR) (Panavia V5) or provisional cement (PR) (RelyX Temp NE), and screw-retained one-piece titanium group (TI). Removal torques were measured in Newton-centimeter (Ncm) before and after 500,000-cycle cyclic loading with forces ranging from 20 to 200 N at 15 Hz. The RTL percentage in each group was calculated. The paired -test was used to detect the difference between pre-loading (RT1) and post-loading removal torque (RT2) in each group and 1-way ANOVA was used to detect the difference of RTL percentage between groups.

Results: The post-loading removal torques in all groups were significantly lower than their pre-loading removal torques ( < 0.001). The 1-way ANOVA test found no significant difference in the RTL% between the study groups. The PR group exhibited the lower RTL% (30.74 ± 7.3%), followed by the TI (30.78 ± 5.6%), AR (32.12 ± 2.5%), and SC (35.71 ± 5.1%) groups.

Conclusion: Combined screw- and cement-retained restorations exhibited similar RTL compared with other restoration designs, and cyclic loading significantly affected the removal torque.

Clinical Significance: Combined screw- and cement-retained restorations can be utilized in single-tooth situations, offering a comparable impact on screw joint stability while providing benefit of retrievability. Cyclic loading significantly influences joint stability, periodic checkup for screw loosening is recommended. How to cite this article: Jongsiri S, Arksornnukit M, Homsiang W, . Effect of Restoration Design on the Removal Torque Loss of Implant-supported Crowns after Cyclic Loading. J Contemp Dent Pract 2023;24(12):951-956.

Download full-text PDF

Source
http://dx.doi.org/10.5005/jp-journals-10024-3604DOI Listing

Publication Analysis

Top Keywords

cyclic loading
28
removal torque
16
combined screw-
16
screw- cement-retained
16
restoration design
12
torque loss
12
implant-supported crowns
12
crowns cyclic
12
rtl percentage
12
removal torques
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!