Effects on mechanical power of different devices used for inhaled sedation in a bench model of protective ventilation in ICU.

Ann Intensive Care

Hospices Civils de Lyon, Service de Médecine Intensive -Réanimation, Hôpital Edouard Herriot, 5 Place d'Arsonval, 69437, Lyon Cedex 03, France.

Published: January 2024

Background: Inhaled sedation during invasive mechanical ventilation in patients with acute respiratory distress syndrome (ARDS) has received increasing attention. However, inhaled sedation devices increase dead-space ventilation and an undesirable effect is the increase in minute ventilation needed to maintain CO removal. A consequence of raising minute ventilation is an increase in mechanical power (MP) that can promote lung injury. However, the effect of inhaled sedation devices on MP remains unknown.

Methods: We conducted a bench study to assess and compare the effects of three devices delivering inhaled sevoflurane currently available in ICU (AnaConDa-50 mL (ANA-50), AnaConDa-100 mL (ANA-100), and MIRUS) on MP by using a test lung model set with three compliances (20, 40, and 60 mL/cmHO). We simulated lung-protective ventilation using a low tidal volume and two levels of positive end-expiratory pressure (5 and 15 cmHO) under ambient temperature and dry conditions. Following the insertion of the devices, either the respiratory rate or tidal volume was increased in 15%-steps until end-tidal CO (EtCO) returned to the baseline value. MP was calculated at baseline and after EtCO correction using a simplified equation.

Results: Following device insertion, the EtCO increase was significantly greater with MIRUS (+ 78 ± 13%) and ANA-100 (+ 100 ± 11%) than with ANA-50 (+ 49 ± 7%). After normalizing EtCO by adjusting minute ventilation, MP significantly increased by more than 50% with all inhaled sedation devices compared to controls. The lowest increase in MP was observed with ANA-50 (p < 0.05 versus ANA-100 and MIRUS). The Costa index, another parameter assessing the mechanical energy delivered to the lungs, calculated as driving pressure × 4 + respiratory rate, significantly increased by more than 20% in all experimental conditions. Additional experiments performed under body temperature, ambient pressure, and gas saturated with water vapor conditions, confirmed the main results with an increase in MP > 50% with all devices after normalizing EtCO by adjusting minute ventilation.

Conclusion: Inhaled sedation devices substantially increased MP in this bench model of protective ventilation, which might limit their benefits in ARDS.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10825094PMC
http://dx.doi.org/10.1186/s13613-024-01245-xDOI Listing

Publication Analysis

Top Keywords

inhaled sedation
24
sedation devices
16
minute ventilation
12
mechanical power
8
devices
8
bench model
8
model protective
8
ventilation
8
protective ventilation
8
tidal volume
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!