Safety and feasibility study of a novel robotic system in an in vivo porcine vascular model.

CVIR Endovasc

2nd Department of Radiology, Interventional Radiology Unit, Medical School, National and Kapodistrian University of Athens, Attikon" University General Hospital, 1st Rimini St, Chaidari, 12461, Athens, GR, Greece.

Published: January 2024

Purpose: The goal of this preclinical study is to assess the functionality, technical feasibility, and safety of a new vascular robotic LIBERTY 3 System, in the microcatheterization of vascular targets using a range of guidewires and microcatheters.

Material And Methods: An anesthetized pig served as an arterial model for the robotic device (LIBERTY3; Microbot Medical Ltd, Yoqneam, IL). The primary efficacy endpoint was the evaluation of its capability to selectively catheterize predetermined distal arterial branches in the liver, kidneys, and mesenteric arteries (technical success), under fluoroscopy guidance. The primary safety endpoint was the occurrence of angiographic acute catheterization-related complications (dissection, thrombosis, embolism, perforation). The catheterizations were conducted by two interventional radiologists that present different work experience in endovascular procedures (18 and 2 years respectively), using a variety of microcatheters and wires. Various procedural parameters such as functionality, practicality, ease of use, and time required for selective catheterization, were evaluated, and recorded.

Results: All pre-determined arteries were successfully selectively catheterized (100% technical success), by both operators. No angiographic acute complications occurred. The microcatheters and wires were manipulated using the remote portable console in an effortless manner that maintained a high level of accuracy. Mean time for selective catheterization was 131 ± 82 s. The robot's conversion function to manual operation was successfully demonstrated.

Conclusion: Robotic navigation and catheterization of selected target arteries were accomplished without observable vascular damage, suggesting that the LIBERTY 3 robotic system is a reliable and safe tool for robotic-assisted endovascular navigation. Further experimental studies are required to evaluate safety and efficacy prior to introduction into clinical practice.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10821852PMC
http://dx.doi.org/10.1186/s42155-024-00425-xDOI Listing

Publication Analysis

Top Keywords

robotic system
8
technical success
8
angiographic acute
8
microcatheters wires
8
selective catheterization
8
robotic
5
safety
4
safety feasibility
4
feasibility study
4
study novel
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!