Backscatter interferometry (BSI) is a refractive index (RI) detection method that is easily integrated with capillary electrophoresis (CE) and is capable of detecting species ranging from inorganic ions to proteins without additional labels or contrast agents. The BSI signal changes linearly with the square of the separation voltage which has been used to quantify sample injection, but has not been explored as a potential signal enhancement mechanism in CE. Here we develop a mathematical model that predicts a signal enhancement at high field strengths, where the BSI signal is dominated by the voltage dependent mechanism. This is confirmed in both simulation and experiment, which show that the analyte peak area grows linearly with separation voltage at high field strengths. This effect can be exploited by adjusting the background electrolyte (BGE) to increase the conductivity difference between the BGE and analyte zones, which is shown to improve BSI performance. We also show that this approach has utility in small bore capillaries where larger separation fields can be applied before excess Joule heating degrades the separation. Unlike other optical detection methods that generally degrade as the optical pathlength is reduced, the BSI signal-to-noise can improve in small bore capillaries as the larger separation fields enhance the signal.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10808210 | PMC |
http://dx.doi.org/10.1038/s41598-024-52621-3 | DOI Listing |
Talanta
December 2024
Department of Chemistry, Simon Fraser University, Burnaby, British Columbia, V5A 1S6, Canada. Electronic address:
A microanalytical technique based on the photothermal effect in conjunction with back-scattering interferometry (BSI) using a single laser beam was developed for quantitative detection of heavy metals. After the chromogenic reaction of an analyte in a capillary tube, the photothermal effect induced by irradiation with the same laser beam leads to a change of the refractive index of the solution, which can be "quantified" using the BSI technique. For prove-of-concept, Cu(II) was chosen as the trial analyte, for which the solution changes to purplish through reacting with the chromogenic reagent; a single laser beam of 532 nm was adapted for both inducing the photothermal effect and realizing BSI detection.
View Article and Find Full Text PDFPhys Rev E
October 2024
Department of Chemistry and Physics, Mount Royal University, Calgary, Alberta, Canada.
In data mining, density-based clustering, which entails classifying datapoints according to their distributions in some space, is an essential method to extract information from large datasets. With the advent of software-based radio, ionospheric radars are capable of producing unprecedentedly large datasets of plasma turbulence backscatter observations, and new automatic techniques are needed to sift through them. We present an algorithm to automatically identify and track clusters of radar echoes through time, using dbscan, a celebrated density-based clustering method for noisy point clouds.
View Article and Find Full Text PDFAnal Chem
August 2024
Ralph N. Adams Institute for Bioanalytical Chemistry, Department of Chemistry, University of Kansas, 2030 Becker Drive, Lawrence, Kansas 66047, United States.
Refractive index (RI) detection using backscatter interferometry (BSI) enables universal detection in capillary electrophoresis (CE). BSI detection is a versatile on-capillary approach that is easily integrated with capillary or microfluidic channels, straightforward to miniaturize, and inexpensive. The focused BSI light source can also double as the excitation source for fluorescence, enabling simultaneous universal (BSI) and specific (fluorescence) signals from the same detection volume.
View Article and Find Full Text PDFJ Biomed Opt
June 2024
University of Illinois Urbana-Champaign, Beckman Institute for Advanced Science and Technology, Urbana, Illinois, United States.
Significance: Full-field optical coherence microscopy (FF-OCM) is a prevalent technique for backscattering and phase imaging with epi-detection. Traditional methods have two limitations: suboptimal utilization of functional information about the sample and complicated optical design with several moving parts for phase contrast.
Aim: We report an OCM setup capable of generating dynamic intensity, phase, and pseudo-spectroscopic contrast with single-shot full-field video-rate imaging called bichromatic tetraphasic (BiTe) full-field OCM with no moving parts.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!