Purpose: To evaluate the clinical value of CT-guided radiofrequency ablation (RFA) in the diagnosis and treatment of pulmonary metastases under optical and electromagnetic navigation.

Methods: Data on CT-guided radiofrequency ablation treatment of 93 metastatic lung lesions in 70 patients were retrospectively analyzed. There were 46 males and 24 females with a median age of 60.0 years (16-85 years). All lesions were ≤3cm in diameter. 57 patients were treated with 17 G radiofrequency ablation needle puncture directly ablated the lesion without biopsy, and 13 patients were treated with 16 G coaxial needle biopsy followed by radiofrequency ablation. There were 25 cases in the optical navigation group, 25 in the electromagnetic navigation group, and 20 in the non-navigation group. The navigation group was performed by primary interventionalists with less than 5 years of experience, and the non-navigation group was performed by interventionalists with more than 5 years of experience.

Result: All operations were successfully performed. There was no statistically significant difference in the overall distribution of follow-up results among the optical, electromagnetic, and no navigation groups. Complete ablation was achieved in 84 lesions (90.3%). 7 lesions showed incomplete ablation and were completely inactivated after repeat ablation. 2 lesions progressed locally, and one of them still had an increasing trend after repeat ablation. No serious complications occurred after the operation.

Conclusions: Treatment with optical and electromagnetic navigation systems by less experienced operators has similar outcomes to traditional treatments without navigational systems performed by more experienced operators.

Download full-text PDF

Source
http://dx.doi.org/10.1080/02656736.2023.2300333DOI Listing

Publication Analysis

Top Keywords

radiofrequency ablation
20
optical electromagnetic
16
electromagnetic navigation
16
ct-guided radiofrequency
12
navigation group
12
ablation
9
patients treated
8
non-navigation group
8
group performed
8
interventionalists 5 years
8

Similar Publications

Background: Pulsed field ablation (PFA) is gaining recognition as a nonthermal, tissue-specific technique for the treatment of atrial fibrillation (AF). The preclinical evaluation of the investigated novel PFA system from Insight Medtech Co. Ltd has demonstrated feasibility, safety, and 30-day efficacy for pulmonary vein isolation (PVI) in the swine model.

View Article and Find Full Text PDF

Evaluation of Local Tumor Outcomes Following Microwave Ablation of Colorectal Liver Metastases Using CT Imaging: A Comparison of Visual versus Quantitative Methods.

Radiol Imaging Cancer

January 2025

From the Department of Radiology, Royal Marsden Hospital NHS Foundation Trust, 203 Fulham Road, London SW3 6JJ, England (J.D.S., L.K., L.P., J.M., N.K., D.M.K., E.J.); Institute of Cancer Research, London, England (N.P., D.M.K.); and Department of Radiology and Nuclear Medicine, Rijnstate Hospital, Arnhem, the Netherlands (W.O.).

Purpose To compare visual versus quantitative ablation confirmation for identifying local tumor progression and residual tumor following microwave ablation (MWA) of colorectal liver metastases (CRLM). Materials and Methods This retrospective study included patients undergoing MWA of CRLM from October 2014 to February 2018. Two independent readers visually assessed pre- and postprocedure images and semiquantitatively scored for incomplete ablation, using a six-point Likert scale, and extracted quantitative imaging metrics of minimal ablative margin (MAM) and percentage of tumor outside of the ablation zone, using both rigid and deformable registration.

View Article and Find Full Text PDF

This study evaluates the feasibility of Apple Vision Pro goggles as an augmented reality (AR) surgical navigation tool for laparoscopic-assisted ultrasound-guided radiofrequency ablation (RFA) of liver tumors. Traditional RFA is effective but challenging due to the integration of multiple imaging modalities. The primary aim of this research is to assess how Vision Pro goggles can enhance the surgical navigation process during RFA, improving tumor localization and the overall effectiveness of the procedure.

View Article and Find Full Text PDF

Pulsed field ablation (PFA) is a catheter-based procedure that utilizes short high voltage and short-duration electrical field pulses to induce tissue injury. The last decade has yielded significant scientific progress and quickened interest in PFA as an energy modality leading to the emergence of the clinical use of PFA technologies for the treatment of atrial fibrillation. It is generally agreed that more research is needed to improve our biophysical understanding of PFA for clinical cardiac applications as well as its potential as a potential alternative energy source to thermal ablation modalities for the treatment of other arrhythmias.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!