Statement Of Problem: Implant-supported rehabilitations are an increasingly frequent practice to replace lost teeth. Before clinical application, all implant components should demonstrate suitable durability in laboratory studies, through fatigue tests.

Objective: The purpose of this in vitro study was to evaluate the integrity and wear of implant components using SEM, and to assess the axial displacement of the implant-abutment assembly by Micro-CT, in different implant connections, after three distinct mechanical requests.

Materials And Methods: Four KLOCKNER implants (external connection SK2 and KL; and internal connection VEGA and ESSENTIAL) were submitted to three different mechanical requests: single tightening, multiple tightening, and multiple tightening and cyclic loading (500 N × 100 cycles). A total of 16 samples were evaluated by SEM, by the X-ray Bragg-Brentano method to obtain residual stresses, and scratch tests were realized for each surface and Micro-CT (4 control samples; 4 single tightening; 4 multiple tightening; 4 multiple tightening and cyclic loading). All dental implants were fabricated with commercially pure titanium (grade 3 titanium). Surface topography and axial displacement of abutment into the implant, from each group, were evaluated by SEM and Micro-CT.

Results: In the manufacturing state, implants and abutments revealed minor structural changes and minimal damage from the machining process. The application of the tightening torque and loading was decisive in the appearance and increase in contact marks on the faces of the hexagon of the abutment and the implant. Vega has the maximum compressive residual stress and, as a consequence, higher scratch force. The abutment-implant distances in SK2 and KL samples did not show statistically significant differences, for any of the mechanical demands analyzed. In contrast, statistically significant differences were observed in abutment-implant distance in the internal connection implants Vega and Essential.

Conclusions: The application of mechanical compression loads caused deformation and contact marks in all models tested. Only internal connection implants revealed an axial displacement of the abutment into the implant, but at a general level, a clear intrusion of the abutment into the implant could only be confirmed in the Essential model, which obtained its maximal axial displacement with cyclic loading.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10820107PMC
http://dx.doi.org/10.3390/ma17020434DOI Listing

Publication Analysis

Top Keywords

axial displacement
16
tightening multiple
16
multiple tightening
16
abutment implant
16
internal connection
12
cyclic loading
12
dental implants
8
three mechanical
8
vitro study
8
implant components
8

Similar Publications

Scaffolds' designs and physical properties have an important place in tissue engineering. Using different biomaterials, scaffolds with other structures can be developed. The thermal and mechanical properties of biomaterials used in producing scaffolds with the fused deposition modeling method are significant for the application's success.

View Article and Find Full Text PDF

Preliminary validation for an online configuration determination method of a thin film buckling under point contact force.

Sci Rep

December 2024

Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, South Korea.

The authors previously developed an online thin film buckling configuration determination method for a mini basket type mapping catheter prototype, which incorporates eight thin film sensor strips. In the prior study, no external force was applied to the thin film, and only axial displacement was adjusted to induce buckling in the thin film. Extending this prior work, a preliminary methodological validation is conducted for an online configuration determination method of thin film buckling under a point contact force.

View Article and Find Full Text PDF

The prevalence of radiographic trochlear dysplasia in patients with patellar fractures.

BMC Musculoskelet Disord

December 2024

Division of Orthopaedic and Trauma Surgery, University Hospitals of Geneva, 4 Rue Gabrielle-Perret-Gentil, Geneva, CH-1205, Switzerland.

Purpose: Trochlear dysplasia is found in 3.2% (95% confidence interval (CI) 1.2-6.

View Article and Find Full Text PDF

Introduction Spinal fusion surgery with pedicle screws is commonly performed to stabilize the spine of osteoporotic patients. However, securing a strong screw fixation in osteoporotic bone presents significant challenges due to the reduced bone density. This study aimed to compare the biomechanical performance in an osteoporotic bone model of pedicle screws inserted using two different techniques, the Jamshidi needle technique and the pedicle probe technique, as well as the influence of tapping on both these techniques.

View Article and Find Full Text PDF

Background: Hemophilic arthritis (HA) is associated with significant changes in the morphology of mature knee joints due to abnormal growth plate development. Previous studies have established marked distinctions between the femur and tibia of subjects with Haemophilia and those with osteoarthritis (OA). This study explored the morphological characteristics of the patella and patellofemoral joint in subjects with Haemophilia.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!