Real-time in-line broad protein content monitoring in biomanufacturing downstream unit operations enables the ability to optimize and afford consistent protein recovery. Protein determination from 2 to 400 mg/mL is demonstrated herein via real-time dual-wavelength LED photometric sensor configured at 280 and 310 nm. The figures of merit of this approach include measurement accuracy within the common acceptance criteria of 100 % ± 5 with negligible bias across the linear dynamic ranges. This work expands the utility of an LED based photometric sensor for biopharmaceutical process analytical technology (PAT) applications. It is also congruent with process digitalization and automation industry 4.0 concepts underpinned by Quality by Design (QbD) principles.