Time-of-flight scatter rejection in x-ray radiography.

Phys Med Biol

Institut Interdisciplinaire d'Innovation Technologique (3IT), Université de Sherbrooke, Sherbrooke, Québec, Canada.

Published: February 2024

Time-of-flight (TOF) scatter rejection allows for identifying and discarding scattered photons without the use of an anti-scatter grid (ASG). Although TOF scatter rejection was initially presented for cone-beam computed tomography, we propose, herein, to extend this approach to x-ray radiography. This work aims to evaluate with simulations if TOF scatter rejection can outperform ASGs for radiography.GATE was used to simulate the radiography of a head and a torso and a water cylinder with bone inserts in a system with total timing jitters from 0 ps up to 500 ps full-width-at-half-maximum. The transmission factor of TOF scatter rejection for primary and scattered photons was evaluated as if it were a virtual ASG.With a total timing jitter of 50 ps, TOF scatter rejection can reach a selectivity of 4.93 with a primary photons transmission of 99%. Reducing the timing jitter close to 0 ps increases the selectivity up to 15.85 for a head and torso radiography, outperforming typical ASGs which usually have a selectivity from 2.5 to 10 with a primary photons transmission from 50% to 70%.This suggests that TOF scatter rejection may be suitable to replace ASGs in applications requiring lower radiation exposure if sufficiently low timing jitter is achieved.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/ad1f85DOI Listing

Publication Analysis

Top Keywords

scatter rejection
28
tof scatter
24
timing jitter
12
x-ray radiography
8
scattered photons
8
head torso
8
total timing
8
primary photons
8
photons transmission
8
rejection
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!