Clinical Relevance: Clinical optical coherence tomography devices are widely used in optometry and ophthalmology and may be used to measure contact lens base curvature radius and visualise contact lens edge shape.
Background: Knowledge of contact lens geometry facilitates fitting, while optical coherence tomography provides a powerful means of measuring geometrical form. This study evaluates the performance of a clinical optical coherence tomography device (3D OCT-1000) in characterising contact lens edge shape and measuring the back optic zone radius of rigid gas-permeable contact lenses in vitro.
Methods: First, an opto-mechanical optical coherence tomography contact lens adaptor was designed and 3D-printed to facilitate a contact lens being imaged using a commercial optical coherence tomography device. Second, several image-processing algorithms and a simple calibration method were developed to measure the back optic zone radius in optical coherence tomography B-scans. Finally, based on the findings of two experiments, B-scan performance was evaluated in terms of 1) capacity to differentiate between contact lens edge geometries, and 2) capacity to obtain accurate and repeatable back optic zone radius measurements. Statistical and graphical analyses were performed to characterise reliability and reproducibility.
Results: The 3D OCT-1000 and adaptor combination was capable of acquiring images of sufficient quality to discriminate between soft and rigid contact lens edge geometries. Additionally, statistical analysis of the rigid contact lens measurements demonstrated satisfactory back optic zone radius measurement accuracy and reproducibility.
Conclusion: This study demonstrates that a 3D OCT-1000 fitted with an opto-mechanical adaptor combination can be used to assess contact lens edges in vitro and that this clinical optical coherence tomography device, combined with image processing and linear calibration of the B-scans, is capable of obtaining back optic zone radius measurements of rigid gas-permeable contact lenses that are close to the ISO 18,369-2:2018 manufacturing tolerance range (±0.05 mm).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/08164622.2024.2302832 | DOI Listing |
Clin Exp Optom
January 2025
Eurolens Research, Division of Pharmacy and Optometry, The University of Manchester, Manchester, UK.
Clinical Relevance: Knowledge of contact lens prescribing trends can (a) assist practitioners to benchmark their own prescribing habits, (b) help the contact lens industry understand preferred products, and (c) support academics in scholarly writings.
Background: This work aims to document contact lens prescribing trends in Australia over the past quarter of a century.
Methods: An annual survey of contact lens prescribing trends was conducted in Australia each year from 2000 to 2024, inclusive, by asking optometrists to provide information relating to 10 consecutive contact lens fits undertaken between January and March.
J Control Release
January 2025
Oujiang Laboratory (Zhejiang Lab for Regenerative Medicine, Vision and Brain Health), Cixi Biomedical Research Institute, School of Pharmaceutical Sciences, Wenzhou Medical University, Ningbo, China. Electronic address:
Severe corneal injuries can cause visual impairment even blindness. Surgically stitching or implanting biomaterials have been developed, but their implementation requires professional surgeons, failing to address the immediate need of medical treatment. The pressing challenge lies in developing multifunctional biomaterials that enable self-management of corneal injuries.
View Article and Find Full Text PDFCont Lens Anterior Eye
January 2025
State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, Guangzhou, China. Electronic address:
Purpose: To develop a Chinese version of the 8-item Contact Lens Dry Eye Questionnaire (CLDEQ-8) with cross-cultural adaption and clinical validation among soft contact lens wearers (SCL) in China.
Methods: The translation and adaptation of the Chinese CLDEQ-8 (C-CLDEQ-8) followed a rigorous methodological process based on cross-cultural research guidelines consisting of three main phases: 1) forward and backward translation of original CLDEQ-8 into Chinese, 2) cross-cultural adaptation performed by medical advising committee, and 3) clinical validation and repeatability test of the translated questionnaire on Chinese SCL wearers. Rasch analysis was also performed to investigate the psychometric properties of C-CLDEQ-8.
Am J Ophthalmol
January 2025
Department of Ophthalmology, University of Warmia and Mazury, Olsztyn, Poland; Institute for Research in Ophthalmology, Foundation for Ophthalmology Development, Poznan Poland. Electronic address:
Purpose: Treatment of myopia has been informed by more than 3 decades of clinical trials and other observations. However, controversies regarding myopia control remain, such as when to stop treatment and what is the long-term efficacy of treatment. This perspective aims to describe clinically relevant and current controversies regarding myopia treatment.
View Article and Find Full Text PDFOphthalmic Physiol Opt
January 2025
Elite School of Optometry, Medical Research Foundation, Chennai, India.
Purpose: To investigate changes in fluid reservoir turbidity parameters over time and its influence on visual performance in eyes with ocular surface disorders (OSD) wearing scleral contact lenses (SL).
Methods: Thirteen eyes with OSD were assessed for corrected distance visual acuity, contrast sensitivity (CS) and fluid reservoir turbidity using anterior segment optical coherence tomography at baseline, after 5 min and 0.5, 1, 2, 3 and 4 h of SL wear on day 1 and after 1 month.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!