Background: This study aims to present the feasibility of developing a synchrotron-based proton ultra-high dose rate (UHDR) pencil beam scanning (PBS) system.

Methods: The RF extraction power in the synchrotron system was increased to generate 142.4 MeV pulsed proton beams for UHDR irradiation at ~100 nA beam current. The charge per spill was measured using a Faraday cup. The spill length and microscopic time structure of each spill was measured with a 2D strip transmission ion chamber. The measured UHDR beam fluence was used to derive the spot dwell time for pencil beam scanning. Absolute dose distributions at various depths and spot spacings were measured using Gafchromic films in a solid-water phantom.

Results: For proton UHDR beams at 142.4 MeV, the maximum charge per spill is 4.96 ± 0.10 nC with a maximum spill length of 50 ms. This translates to an average beam current of approximately 100 nA during each spill. Using a 2 × 2 spot delivery pattern, the delivered dose per spill at 5 cm and 13.5 cm depth is 36.3 Gy (726.3 Gy/s) and 56.2 Gy (1124.0 Gy/s), respectively.

Conclusions: The synchrotron-based proton therapy system has the capability to deliver pulsed proton UHDR PBS beams. The maximum deliverable dose and field size per pulse are limited by the spill length and extraction charge.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10778151PMC
http://dx.doi.org/10.3390/cancers16010221DOI Listing

Publication Analysis

Top Keywords

pencil beam
12
beam scanning
12
spill length
12
ultra-high dose
8
dose rate
8
rate uhdr
8
synchrotron-based proton
8
1424 mev
8
pulsed proton
8
beam current
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!