Background: the use of handheld ultrasonography (HHUS) devices is well established in prehospital emergency diagnostics, as well as in intensive care settings. This is based on several studies in which HHUS devices were compared to conventional high-end ultrasonography (HEUS) devices. Nonetheless, there is limited evidence regarding potential variations in B-scan quality among HHUS devices from various manufacturers, and regarding whether any such differences hold clinical significance in intensive care medicine settings.

Methods: this study included the evaluation of eight HHUS devices sourced from diverse manufacturers. Ultrasound videos of five previously defined sonographic questions (volume status/inferior vena cava, pleural effusion, pulmonary B-lines, gallbladder, and needle tracking in situ) were recorded with all devices. The analogue recording of the same pathologies with a HEUS device served as gold standard. The corresponding findings (HHUS and HEUS) were then played side by side and evaluated by sixteen intensive care physicians experienced in sonography. The B-scan quality and the clinical significance of the HHUS were assessed using a five-point Likert scale (5 points = very good; 1 point = insufficient).

Results: both in assessing the quality of B-scans and in their ability to answer clinical questions, the HHUS achieved convincing results-regardless of the manufacturer. For example, only 8.6% (B-scan quality) and 9.8% (clinical question) of all submitted assessments received an "insufficient" rating. One HHUS device showed a significantly higher ( < 0.01) average points score in the assessment of B-scan quality (3.9 ± 0.65 points) and in the evaluation of clinical significance (4.03 ± 0.73 points), compared to the other devices.

Conclusions: HHUS systems are able to reliably answer various clinical intensive care questions and are-while bearing their limitations in mind-an acceptable alternative to conventional HEUS devices. Irrespective of this, the present study was able to demonstrate relevant differences in the B-scan quality of HHUS devices from different manufacturers.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10795594PMC
http://dx.doi.org/10.3390/diagnostics14010054DOI Listing

Publication Analysis

Top Keywords

b-scan quality
24
intensive care
20
hhus devices
20
clinical significance
16
hhus
11
quality clinical
8
significance intensive
8
devices
8
heus devices
8
quality hhus
8

Similar Publications

Endoscopic optical coherence tomography (OCT) is widely used in the detection of morphological alterations in luminal organs, which provides high-resolution, three-dimensional (3D) images of internal tissues. In most cases, lesions are revealed early by microvascular pathological changes in cavity tissues. There is a significant demand for the performance of endoscopic OCT angiography (OCTA) to visualize the superficial capillaries.

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the intraretinal hyperreflective line (IHL), a new finding seen through optical coherence tomography (OCT) that suggests damage in the retina's layers due to various conditions.
  • Through a retrospective analysis of records from 2019 to 2023, researchers identified IHLs in 40 eyes of 38 participants, noting its connection to various retinal diseases and prior surgeries.
  • Key findings included the presence of IHL in 35% of eyes with prior vitreoretinal surgery, as well as associations with conditions like full-thickness macular holes, diabetic macular edema, and idiopathic macular telangiectasia.
View Article and Find Full Text PDF
Article Synopsis
  • The study compares structural optical coherence tomography (OCT) and infrared reflectance (IR) imaging to analyze differences in non-exudative macular neovascularizations (NE MNVs) caused by age-related macular degeneration (AMD) and those linked to the pachychoroid spectrum.
  • It assessed patients with documented NE-MNVs, focusing on treatment-naïve cases that remained non-exudative for at least a year and had high-quality imaging available.
  • Results indicated that NE-MNVs in pachychoroid eyes exhibited sharper, more inhomogeneous pigment epithelium detachments and were more likely to show a hyporeflective halo than those in AMD eyes.
View Article and Find Full Text PDF

AMD-SD: An Optical Coherence Tomography Image Dataset for wet AMD Lesions Segmentation.

Sci Data

September 2024

Ophthalmic Center, The Second Affiliated Hospital, Jiangxi Medical College, Nanchang University, Nanchang, 330000, P. R. China.

Wet Age-related Macular Degeneration (wet AMD) is a common ophthalmic disease that significantly impacts patients' vision. Optical coherence tomography (OCT) examination has been widely utilized for diagnosing, treating, and monitoring wet AMD due to its cost-effectiveness, non-invasiveness, and repeatability, positioning it as the most valuable tool for diagnosis and tracking. OCT can provide clear visualization of retinal layers and precise segmentation of lesion areas, facilitating the identification and quantitative analysis of abnormalities.

View Article and Find Full Text PDF

(1) Background: Ocular emergencies account for 1.5-3% of emergency department (ED) visits and require urgent diagnosis to prevent serious complications. Ultrasonography is a crucial, non-invasive diagnostic tool for these conditions but traditionally lacks portability and integration with modern electronic smart devices.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!