Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: There is considerable debate about the hemodynamic effects of veno-arterial extracorporeal membrane oxygenation (VA-ECMO).
Aims: To evaluate the changes in left ventricular (LV) function, volumes, and work in patients treated with VA-ECMO using invasive LV catheterization and three-dimensional echocardiographic volumes.
Methods: Patients on VA-ECMO underwent invasive hemodynamic evaluation due to concerns regarding candidacy for decannulation. Hemodynamic parameters were reported as means±standard deviations or medians (interquartile ranges) after evaluating for normality. Paired comparisons were done to evaluate hemodynamics at the baseline (highest) and lowest tolerated levels of VA-ECMO support.
Results: Twenty patients aged 52.3 ± 15.8 years were included. All patients received VA-ECMO for refractory cardiogenic shock (5/20 SCAI stage D, 15/20 SCAI stage E). At 3.0 (2.0, 4.0) days after VA-ECMO cannulation, the baseline LV ejection fraction was 20% (15%, 27%). The baseline and lowest VA-ECMO flows were 4.0 ± 0.6 and 1.5 ± 0.6 L/min, respectively. Compared to the lowest flow, full VA-ECMO support reduced LV end-diastolic volume [109 ± 81 versus 134 ± 93 mL, p = 0.001], LV end-diastolic pressure (14 ± 9 vs. 19 ± 9 mmHg, p < 0.001), LV stroke work (1858 ± 1413 vs. 2550 ± 1486 mL*mmHg, p = 0.002), and LV pressure-volume area (PVA) (4507 ± 1910 vs. 5193 ± 2388, p = 0.03) respectively. Mean arterial pressure was stable at the highest and lowest flows (80 ± 16 vs. 75 ± 14, respectively; p = 0.08) but arterial elastance was higher at the highest VA-ECMO flow (4.9 ± 2.2 vs lowest flow 2.7 ± 1.6; p < 0.001).
Conclusions: High flow VA-ECMO support significantly reduced LV end-diastolic pressure, end-diastolic volume, stroke work, and PVA compared to minimal support. The Ea was higher and MAP was stable or minimally elevated on high flow.